3
Testing 16860 Performance
8
9
10 Continue increasing the generator frequency until the logic analyzer displays an error that the
11 Without changing the frequency, clear the message by clicking OK. Then click the Run Repetitive
12 If the analyzer displays an error again, lower the frequency by 1MHz and run again. If the test
13 Continue changing the frequency and running unit the frequency can no longer be increased and
14 Set the generator to the highest frequency that did not cause an error.
15 Click Run Repetitive
16 Wait for logic analyzer to complete 100 acquisitions at the new pulse generator frequency
17 Click the Stop
18 Note the generator frequency setting this will be used in the next section to verify the data rate.
64
Click the Run Repetitive
acquiring data repeatedly. Acquired data will start appearing in the Listing window.
Start increasing the frequency on the pulse generator by 1 MHz increments while simultaneously
observing the logic analyzer data acquisition status.
data could not be displayed, or the clock is too fast. It is possible that the error is caused by noise
introduced as the generator frequency is changed.
toolbar button to start a repetitive run again at the same frequency. Running at the same
frequency checks to see if the error message was caused by noise due to the changing of the
generator frequency.
does not display an error, increase the frequency by 1 MHz and run again.
the error message not displayed.
.
without displaying any error. If an error is displayed, decrease the pulse generator frequency by 1
MHz and then again wait for 100 acquisitions at this new frequency without any error. Repeat this
step until you get 100 acquisitions without any error display.
toolbar button to stop the data acquisition.
toolbar button to start a repetitive run on the logic analyzer for