CB-7080 & CB-7080D Counter/Timer User's Guide
$AA1L(data)
Description
: Set the low trigger level of non-isolated input. Refer to
setting" for more information.
Syntax
: $AA1L(data)[chk](cr)
$
AA
(data)
[chk]
(cr)
Response
:
Valid response
Invalid response
no response
!
?
AA
[chk]
(cr)
Examples
:
Low trigger level=0.8 volt:
command:
response:
Low trigger level=1.0 volt:
command:
response:
Note:
The default value for the low trigger level is 0.8V.
A delimiter character
2-character HEX module address, from 00 to FF
2-character decimal value for low trigger level. The unit is 0.1V and the
range can be from 0.0 to 5.0V
2-character checksum. If checksum is disabled, no [chk].
0x0D
!AA[chk](cr)
?AA[chk](cr)
Syntax error, communication error or address error
A delimiter character indicating a valid response
A delimiter character indicating an invalid response
2-character HEX module address
2-character checksum. If checksum is disabled, no [chk].
0x0D
$011L08(cr)
!01(cr)
$021L10(cr)
!02(cr)
"Programmable threshold voltage