Indicator Status; Remote Commands And Error Codes - Ametek JOFRA DTI050 A Reference Manual

Digital temperature indicator
Table of Contents

Advertisement

5.4.4

Indicator status

Error queue
If an error occurs due to invalid input or buffer overflow, its error code is sent
to the error queue. The error code can be read from the queue with the
command FAULT?. The error queue holds 15 error codes. When it is empty,
FAULT? returns 0. The error queue is cleared when power is reset or when
the clear command *CLS is entered.
Input buffer
Indicator stores all received data in the input buffer. The buffer holds 250
characters. The characters are processed on a first in, first out basis.
5.5

Remote commands and error codes

The following tables list all commands, and their descriptions, that are
accepted by the indicator.
Common commands (table 3)
Command
*CLS
*IDN?
*RST
DTI050 commands (table 4)
Command
FAULT?
FUNC
FUNC?
LOCAL
125821 00
Description
(Clear status.) Clears the error queue.
Identification query. Returns the
manufacturer, model number, and
firmware revision level of the indicator.
Resets the indicator to the power up state.
Description
Returns a fault code.
Sets the measurement function of the DTI050.
Returns the state of the measurement function
of the DTI050.
Sets the DTI050 to local operation.