Dti050 Commands - Ametek JOFRA DTI050 A Reference Manual

Digital temperature indicator
Table of Contents

Advertisement

5.6.2

DTI050 commands

FAULT?
Returns the error code number of an error that has occurred. The command
may be entered when the previous command did not do what it was meant to
do.
For example, if an invalid command was sent, FAULT? would return:
110 which is the code number for an unknown command
being received.
Refer to the Error codes table for more information on error code numbers.
FUNC
Sets the DTI050 to the mode indicated in the argument. The following is a list
of valid arguments:
RTD
THERM
OHMS
TC
MV
For example:
To set the DTI050 into RTD mode send FUNC RTD.
FUNC?
Returns the current mode of the DTI050. For example if the DTI050 is set to
RTD, FUNC? Would return:
RTD
LOCAL
Restores the DTI050 to local operation if it was in remote mode. Also clears
LOCKOUT if the DTI050 was in lockout mode.
125821 00
31-08-2006
31

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Jofra dti050 b

Table of Contents