Character Processing; Response Data Types - Ametek JOFRA DTI050 A Reference Manual

Digital temperature indicator
Table of Contents

Advertisement

5.4.2

Character processing

The data entered into the indicator is processed as follows:
ASCII characters are discarded if their decimal equivalent is less than 32
(space), except 10 (LF) and 13 (CR):
Data is taken as 7-bit ASCII.
The most significant data bit is ignored.
Upper or lower case is acceptable.
5.4.3

Response data types

The data returned by the indicator can be divided into four types:
Integer
For most computers and controllers they are decimal numbers ranging from
-32768 to 32768. For example:
Refer to the Error codes table for more information on error codes.
Floating
Floating numbers have up to 15 significant figures and exponents. For
example:
Character response data (CRD)
Data returned as keywords. For example:
Indefinite ASCII (IAD)
Any ASCII characters followed by a terminator. For example:
26
FAULT? could return 110
P_CVD_A? returns 3.908300E-03
RTD_TYPE? returns PT385_100
*IDN? returns AMETEK, DTI050, 1234567, 1.00
31-08-2006
125821
00