Ametek JOFRA DTI050 A Reference Manual page 32

Digital temperature indicator
Table of Contents

Advertisement

LOCKOUT
Sending this command sets the lockout state, when the DTI050 is in
REMOTE or goes to remote it prohibits use of the keypad completely.
The lockout state can only be cleared by sending the LOCAL
command.
MV?
Returns the raw voltage value from a thermocouple.
For example:
If when measuring a type J at 0 degrees cel sending MV? would
return 0.000000E+00, MV .
OHMS?
Returns the raw Ohm value from a RTD.
For example:
If when measuring a P100-385 at 0 degrees cel sending OHMS?
would return 1.000000E+02, OHM .
REMOTE
Puts the DTI050 in remote mode. From the remote mode the user can still
use the keypad to get back to local unless the command LOCKOUT was
entered before REMOTE. Then the keypad is totally locked out and the user
has to send the LOCAL command to get back to local operation.
RTD_TYPE
Sets the RTD type. The following is a list of RTD types the way they should
be entered after the command:
P10(90)385
P200(90)385
P1K(90)385
P100(90)392
M100(90)428
For example:
RTD_TYPE P100(90)385 sets RTD type to PT100, α=385, ITS-90
32
P50(90)385
P100(90)385
P400(90)385 P500(90)385
P50(90)391
P100(90)391
M10(90)427 M50(90)428
H120(90)672 P100(90)JIS
31-08-2006
125821
00