Load Ranger Wheel Weigh Pad System
5.0
Calibration
This section provides the procedures necessary to calibrate the Load Ranger wheel weigh pads through the built-in displays
and with the Ai-1 indicator.
When using the Ai-1 indicator with the weigh pads, skip to
through the Ai-1 indicator.
Note
5.1
Calibrate with Pad Display
This section describes procedures for setting calibration parameters and the calibration of the Load Ranger Wheel Weigh Pads
through the built-in display.
5.1.1
Complete Calibration Procedure
Follow these steps to set the calibration parameters and complete a calibration:
Press
at any time to back out one level or multiple times to return to the weigh mode.
Note
1. Turn indicator on by pressing
2. Press
. displays.
If
displays instead of
Note
displays. Now continue with
3. Press
. displays.
4. Press
. The current decimal setting displays.
5. Press
to scroll through decimal settings and press
6. Press
. The current division setting displays.
7. Press
to scroll through division settings and press
8. Press
. The current unit setting displays.
9. Press
to scroll through unit settings and press
10. Press
. Range 1 value displays. Use the numeric entry procedure to set value.
For single-range scales, Range 1 is set as the max scale capacity.
Note
11. Press
. displays.
Range 2 should only be set for multi-range scales. For single-range scales, skip to
Note
12. Press
. Range 2 value displays. Use the numeric entry procedure to set value.
13. Press
. displays.
14. Press
. displays.
15. Press
. displays. Make sure weigh pad is unloaded.
16. Press
. briefly displays once the procedure is complete, then displays.
17. Press
. The current setting for the number of calibration points displays.
18. Press
to scroll through settings (1-3) and press
19. Use the numeric entry procedure to set the calibration weight value and press
34
Visit our website
and press
during startup. displays.
, press
, then press
Step
3.
www.RiceLake.com
Section 5.2 on page 35
two times until
to save selection. displays.
to save selection. displays.
to save selection. displays.
to save selection. Calibration point 1 value displays.
. displays.
to calibrate all of the weigh pads
displays. Press
and
Step
14.