Default Factory Calibration; Theoretical Calibration; Calibration Using A Known Weight - Rice Lake SCT-2200 Advanced Series Technical Manual

Hide thumbs Also See for SCT-2200 Advanced Series:
Table of Contents

Advertisement

1.3.1

Default Factory Calibration

The instrument is shipped with the following default calibration settings:
• Capacity – 10,000 kg
• Load cell sensitivity – 2.000 mV/V
• Division – 1
1.3.2

Theoretical Calibration

Use the quick setup menu
1. Press
and
2. Press
and
3. Press
,
and
4. Press
,
and
5. Slowly press
C
• Press
to exit and save the calibration
• Press any other key to exit without saving
1.3.3

Calibration Using a Known Weight

Use the quick setup menu
weight.
1. Select decimals and minimum division using
2. Set capacity using
3. Perform a quick calibration of zero.
4. Set the weight value used for the calibration using
5. Slowly press
C
• Press
to exit and save the calibration
• Press any other key to exit without saving
6
SCT-2200 Weight Transmitter
(Figure 1-3 on page 5
steps 1,2,3, and 4) to perform a theoretical calibration.
to set decimals and minimum division.
to set the total load capacity of the load cells.
to set the mV/V sensitivity of the load cell.
to set the weight value present on the load cell.
multiple times until  displays.
(Figure 1-3 on page 5
steps 1,5,6, and 7) to perform a standard calibration with a known
,
and
.
multiple times until displays.
and
.
,
and
.