Relocate The System For Maintenance - Edwards iH80 Instruction Manual

Ih series. dry pumping system
Table of Contents

Advertisement

A590-00-880 Issue D
Route and secure cables, hoses and pipelines during maintenance to avoid possible risk of trips or
entrapment.
Ensure that any oil or water collected in the secondary containment drip tray is removed before moving the
system.
The enclosure panels should only be removed with the use of a special tool, when the system has been
switched off and allowed to cool sufficiently (as an indication the system should be left for one hour with
cooling water still connected at 1 l min
Wear the appropriate safety-clothing when you come into contact with contaminated components. Dismantle
and clean contaminated components inside a fume-cupboard.
Re-check the pump rotation direction if the electrical supply has been disconnected.
O-ring replacement intervals vary depending on your application.
Dispose of components, grease and oil safely.
Take care to protect sealing-faces from damage.
Do not touch or inhale the thermal breakdown products of fluorinated materials which may be present if the iH
system has been overheated to 260 °C and above. These breakdown products are very dangerous. Fluorinated
materials in the iH system may include oils, greases and seals. The iH system may have overheated if it was misused,
if it malfunctioned or if it was in a fire. Edwards Material Safety Data Sheets for fluorinated materials used in the
pump are available on request: contact your supplier or Edwards. (refer to the address page at the rear of this manual
for contact details).
The iH system requires little user maintenance. Safety sensors fitted to the iH system do not require routine
maintenance. The maintenance operations you can carry out are described in the following sections, any other
maintenance must be carried out by Edwards service centres (refer to
operations depends on your process. Adjust the frequency of maintenance operations according to your experience.
When you maintain the iH system, use Edwards maintenance and service kits. These contain all of the necessary seals
and other components necessary to complete maintenance operations successfully. Ensure that your nitrogen and
cooling-water supplies comply with the data given in
Edwards or refer to the Web.
5.2

Relocate the system for maintenance

The substances that accumulate in the exhaust-pipe, elbow and check-valve can be dangerous. Do
not allow these substances to come into contact with your skin or eyes. Do not inhale vapours from
these substances. Fit blanking caps to the inlet and outlet flanges when you move the exhaust-
pipe, elbow or check-valve around your workplace.
The majority of synthetic oils/grease can cause inflammation of the skin (dermatitis). Safety
precautions must be taken to prevent prolonged skin contact with these substances. Use of
suitable protective gloves and clothing along with a respirator is recommended if contact with the
substance is anticipated.
System process gases and residue can be highly toxic. Take all necessary precautions when
handling components that have, or could have, come into contact with them, including O-rings,
lubricants and all exhaust accessories.
You must use suitable lifting equipment to move the system. It is too heavy to lift by hand.
Page 32
-1
).
Section
Section 2
and that they are connected in parallel. Contact
WARNING
WARNING
© Edwards Limited 2008. All rights reserved.
Edwards and the Edwards logo are trademarks of Edwards Limited.
7). The frequency of maintenance