Appendix 2 B: Errors during recalibration
The following errors are only issued during the recalibration and/or basic calibra-
tion. If one of these errors occurs, the calibration is invalid. The previous response
factors are retained. All of the subsequent measurements are marked as faulty.
These errors cannot be acknowledged and are not reset until a valid recalibration
takes place.
Number
Text
100
Cal:
Retention time
101
Cal:
Response factor
103
Cal:
Total area
Appendix 2 C: Mains failure of the GC 9300
After a mains failure of the analytical computer (Error 02 - Mains failure), a self-test
is carried out by the device after a restart. When this process is completed and the
measuring element is ready, a calibration is initiated automatically (after a speci-
fied wait time). The analysis operation resumes after the calibration.
Description
The newly determined retention times
(current and/or stream values under
"Times") deviate impermissibly from the
specified values ("Computing parame-
ters / Ana.,Cal. limit values")
The newly calculated response factors
show an impermissible deviation ("De-
tail" screen under "Calibration results"
and "Calibration parameters / Cal. limit
values")
The total area determined in the calibra-
tion deviates by more than 30% from the
value of the basic calibration.
Appendix 2: Error messages
Cause
- Pressure / temperature error
during recalibration.
- Calibrating gas supply
- Faulty specification of calibrating
gas concentration
("Detail" screen under "Calibrat-
ing parameters")
- Defective module
- see 100
- see 100
Manual PGC 9300 EN07 June, 18th 2018