RMG PGC 9300 Operating Manual page 39

Measuring element
Hide thumbs Also See for PGC 9300:
Table of Contents

Advertisement

Outlet pipe bypass E1
3
Gas inlets:
A. carrier gas 1
(inlet pressure 5.5 bar)
B. int. calibration gas
(inlet pressure 2-3 bar)
C. ext. test gas
(inlet pressure 2-3 bar)
D. carrier gas 2
(inlet pressure 5.5 bar)
M. sample gases
(inlet pressure 2-3 bar)
A
B
C
1-stream
D
M1
2-stream
M2
3-stream
M3
4-stream
M4
operation
Supply lines for connections A,B,C,D:
1/8" pipe with
clamping ring connection
Supply lines for connections
M1-4:
4 mm
pipe with
clamping ring connection
Supply lines for connections
E1, E2:
12 mm
pipe with
clamping ring connection
Total weight: approx. 75 kg
Figure 6: Gas chromatograph PGC 9300
3 Gas chromatograph PGC 9300
Gas Chromatograph
PGC 9300
Explosion protected design
E2 Outlet pipe sample gas
1
2
4
5
6
400
335
1.) Chromatograph, type CP4900-GC with valve control-
explosion-proof enclosure
2.) Ex(e) connection box
3.) Flow indicator for bypass, sample gas
4.) Gas inlet module from the left with filter plate in
pipe (basic version)
Option:
5.) Inlet filter carrier gas
6.) Membrane filter unit with condensate collectors
(hand valve) for sample gases
215
30
mounting hole
Ø11 x 4
15
295
325
Manual PGC 9300 EN07 June, 18th 2018