RMG PGC 9300 Operating Manual page 54

Measuring element
Hide thumbs Also See for PGC 9300:
Table of Contents

Advertisement

4 Connection and commissioning
Pin assignment
Measuring ele-
ment
50
The cable type NYY-J 3G 4 mm
distances greater than 50 m between the control cabinet and PGC.
Clamping ranges of cable glands (permissible cable diameter):
Manual PGC 9300 EN07 June, 18th 2018
Signal
1
p sample gas +
2
p sample gas -
3
p carrier gas 1 +
4
p carrier gas 1 -
5
TxD+
6
TxD-
7
RxD+
8
RxD-
9
p carrier gas 2 +
10
p carrier gas 2 -
11
int. cal.gas
12
ext. Cal.gas
13
stream 1
14
stream 2
15
stream 3
16
stream 4
17
GND
23
+ 24 V Heater
24
- 24 V Heater
25
+ 24 V
Measuring element
26
- 24 V
Measuring element
number
Ex-PGC
3
3
GC 9300
X5 / 3
X5 / 4
X5 / 1
X5 / 2
X18
(X6 / 3)
(X6 / 4)
X2 / 1
X2 / 3
X1 / 1
X1 / 3
X1 / 5
X1 / 7
X1 / 2, 4, 6, 8
X2 / 2, 4
2
should be used for the power supply (23-26) for
Inner casing
Outer casing
3.1 – 8.6 mm
6.0 – 13.4 mm
6.1 – 11.6 mm
9.5 – 15.9 mm
Recommended cable type
RE-2Y(St) Yv BK
2 x 2 x 0.75 mm²
(for PGC 9304, see below)
Herkulat, 600 A S/FTP
4 x 2 AWG23
RE-2Y(St) Yv BK 4 x 2 x 0.75 mm²
For PGC 9304 (Carrier Gas 2)
Yv BK
RE-2Y(St)
4 x 2 x 0.75 mm²
NYY-J
3G 2.5 mm²
NYY-J
3G 2.5 mm²