RMG PGC 9300 Maintenance Manual

Process gas chromatograph
Hide thumbs Also See for PGC 9300:

Advertisement

Quick Links

Process Gas Chromatograph PGC 9300
MAINTENANCE LOGBOOK
Reliable Measurement of Gas
Status:
2018, JUNE 26TH
Version:
07

Advertisement

Table of Contents
loading

Summary of Contents for RMG PGC 9300

  • Page 1 Process Gas Chromatograph PGC 9300 MAINTENANCE LOGBOOK Reliable Measurement of Gas Status: 2018, JUNE 26TH Version:...
  • Page 2 Contact Manufacturer Our customer service is available for technical queries Address RMG Messtechnik GmbH Otto-Hahn-Straße 5 D-35510 Butzbach +49 6033 897 – 0 Telefon switchboard +49 6033 897 – 0 Telefon service +49 6033 897 – 173 Telefon spare parts +49 6033 897 –...
  • Page 3 Your PGC9300 PGC 930 Maintenace Logbook PGC manufacturer number PGC year of manufacture Meter serial number GC 9300 manufacturer number GC 9300 year of manufacture GC 9300 software version   Calorific value Gas quality Approved for Operator Station Commissioning Type / carrier Column temperature Column pressure...
  • Page 4: Table Of Contents

    Contents Inhalt 1. Instructions for Keeping the Maintenance Logbook ........1 1.1. Inspection....................1 1.1.1. Operating parameters on the GC 9300 ..........1 1.1.2. Cylinder pressures .................. 2 1.2. Maintenance by service technicians ............2 2. Manufacturer's requirements ................ 3 2.1.
  • Page 5: Instructions For Keeping The Maintenance Logbook

    The setpoints for the operating parameters are entered on the first page of this book by the test laboratory during commissioning and shall remain unchanged during operation, unless altered by an RMG service technician. The actual values of the operating parameters can be checked at any time on the GC 9300 analyser under the Status->Meter tab or using RMGViewGC.
  • Page 6: Cylinder Pressures

    1. Instructions for Keeping the Maintenance Logbook 1.1.2. Cylinder pressures The pressure in the gas cylinders is checked directly at the high-pressure gauges of the relevant pressure reducing unit at regular intervals, in other words at every maintenance check, and the information is then documented in the relevant table in Table Section 1 of this logbook.
  • Page 7: Manufacturer's Requirements

    The annual inspection of the PGC 930x must be carried out on the basis of the checklist (Table Section 4) by persons who have received appropriate training from RMG to provide such maintenance. 2.2. Metrological check / (re)calibration Once the maintenance activities described in Section 2 are complete, the device will subsequently undergo routine recalibration.
  • Page 8: Cylinder Pressures

    Table section 1 Table Section 1 Cylinder pressures To be performed at: • Every inspection • Every maintenance Entries made by: • Operator • RMG Service PGC9300 Maintenance Logbook · EN07 · 2018, June 26th...
  • Page 9 Table section 1 / cylinder pressures : Cylinder changed : Device faulty Carriergas I Int. calibration gas Carrier gas II Date Initials    Pressure / bar Pressure / bar Pressure / bar PGC9300 Maintenance Logbook · EN07 · 2018, June 26th...
  • Page 10 Table section 1 / cylinder pressures : Cylinder changed : Device faulty Carriergas I Int. calibration gas Carrier gas II Date Initials    Pressure / bar Pressure / bar Pressure / bar PGC9300 Maintenance Logbook · EN07 · 2018, June 26th...
  • Page 11 Table section 1 / cylinder pressures : Cylinder changed : Device faulty Carriergas I Int. calibration gas Carrier gas II Date Initials    Pressure / bar Pressure / bar Pressure / bar PGC9300 Maintenance Logbook · EN07 · 2018, June 26th...
  • Page 12: Maintenance Measures, General

    Table section 2 Maintenance measures, general To be performed at: • Annual routine maintenance • Software update • Metrological check • (Re)calibration • Repair Entries made by: • Trained personnel • RMG Service PGC9300 Maintenance Logbook · EN07 · 2018, June 26th...
  • Page 13 Table section 2 / maintenance measures, general Maintenance measure Performed by Performed on Measures performed Routine maintenance Table 3 Software update Metrological check Table 4 (Re)calibration Table 4 Defect See below Description of problem Measure Description of problem Measure Signature PGC9300 Maintenance Logbook ·...
  • Page 14 Table section 2 / maintenance measures, general Maintenance measure Performed by Performed on Measures performed Routine maintenance Table 3 Software update Metrological check Table 4 (Re)calibration Table 4 Defect See below Description of problem Measure Description of problem Measure Signature PGC9300 Maintenance Logbook ·...
  • Page 15 Table section 2 / maintenance measures, general Maintenance measure Performed by Performed on Measures performed Routine maintenance Table 3 Software update Metrological check Table 4 (Re)calibration Table 4 Defect See below Description of problem Measure Description of problem Measure Signature PGC9300 Maintenance Logbook ·...
  • Page 16 Table section 2 / maintenance measures, general Maintenance measure Performed by Performed on Measures performed Routine maintenance Table 3 Software update Metrological check Table 4 (Re)calibration Table 4 Defect See below Description of problem Measure Description of problem Measure Signature PGC9300 Maintenance Logbook ·...
  • Page 17 Table section 2 / maintenance measures, general Maintenance measure Performed by Performed on Measures performed Routine maintenance Table 3 Software update Metrological check Table 4 (Re)calibration Table 4 Defect See below Description of problem Measure Description of problem Measure Signature PGC9300 Maintenance Logbook ·...
  • Page 18 Table section 2 / maintenance measures, general Maintenance measure Performed by Performed on Measures performed Routine maintenance Table 3 Software update Metrological check Table 4 (Re)calibration Table 4 Defect See below Description of problem Measure Description of problem Measure Signature PGC9300 Maintenance Logbook ·...
  • Page 19 Table section 2 / maintenance measures, general Maintenance measure Performed by Performed on Measures performed Routine maintenance Table 3 Software update Metrological check Table 4 (Re)calibration Table 4 Defect See below Description of problem Measure Description of problem Measure Signature PGC9300 Maintenance Logbook ·...
  • Page 20 Table section 2 / maintenance measures, general Maintenance measure Performed by Performed on Measures performed Routine maintenance Table 3 Software update Metrological check Table 4 (Re)calibration Table 4 Defect See below Description of problem Measure Description of problem Measure Signature PGC9300 Maintenance Logbook ·...
  • Page 21 Table section 2 / maintenance measures, general Maintenance measure Performed by Performed on Measures performed Routine maintenance Table 3 Software update Metrological check Table 4 (Re)calibration Table 4 Defect See below Description of problem Measure Description of problem Measure Signature PGC9300 Maintenance Logbook ·...
  • Page 22 Table section 2 / maintenance measures, general Maintenance measure Performed by Performed on Measures performed Routine maintenance Table 3 Software update Metrological check Table 4 (Re)calibration Table 4 Defect See below Description of problem Measure Description of problem Measure Signature PGC9300 Maintenance Logbook ·...
  • Page 23: Annual Routine Maintenance

    Table section 3 Table section 3 Annual routine maintenance To be performed at: • Annual routine maintenance Entries made by: • Trained personnel • RMG Service PGC9300 Maintenance Logbook · EN07 · 2018, June 26th...
  • Page 24 Table section 3 / annual routine maintenance Tested: n. OK None Checklist for HP reduction type DRS Functional check of regulator Leak check Functional check of analysis gas heating (if any) Functional check of HP reduction heating (if any) Checklist for cylinder rack Check and documentation of cylinder pressures of carrier gas/gases and calibration gas Functional test of cylinder heating...
  • Page 25 Table section 3 / annual routine maintenance Tested: n. OK None Checklist for HP reduction type DRS Functional check of regulator Leak check Functional check of analysis gas heating (if any) Functional check of HP reduction heating (if any) Checklist for cylinder rack Check and documentation of cylinder pressures of carrier gas/gases and calibration gas Functional test of cylinder heating...
  • Page 26 Table section 3 / annual routine maintenance Tested: n. OK None Checklist for HP reduction type DRS Functional check of regulator Leak check Functional check of analysis gas heating (if any) Functional check of HP reduction heating (if any) Checklist for cylinder rack Check and documentation of cylinder pressures of carrier gas/gases and calibration gas Functional test of cylinder heating...
  • Page 27 Table section 3 / annual routine maintenance Tested: n. OK None Checklist for HP reduction type DRS Functional check of regulator Leak check Functional check of analysis gas heating (if any) Functional check of HP reduction heating (if any) Checklist for cylinder rack Check and documentation of cylinder pressures of carrier gas/gases and calibration gas Functional test of cylinder heating...
  • Page 28 Table section 3 / annual routine maintenance Tested: n. OK None Checklist for HP reduction type DRS Functional check of regulator Leak check Functional check of analysis gas heating (if any) Functional check of HP reduction heating (if any) Checklist for cylinder rack Check and documentation of cylinder pressures of carrier gas/gases and calibration gas Functional test of cylinder heating...
  • Page 29 Table section 3 / annual routine maintenance Tested: n. OK None Checklist for HP reduction type DRS Functional check of regulator Leak check Functional check of analysis gas heating (if any) Functional check of HP reduction heating (if any) Checklist for cylinder rack Check and documentation of cylinder pressures of carrier gas/gases and calibration gas Functional test of cylinder heating...
  • Page 30 Table section 3 / annual routine maintenance Tested: n. OK None Checklist for HP reduction type DRS Functional check of regulator Leak check Functional check of analysis gas heating (if any) Functional check of HP reduction heating (if any) Checklist for cylinder rack Check and documentation of cylinder pressures of carrier gas/gases and calibration gas Functional test of cylinder heating...
  • Page 31 Table section 3 / annual routine maintenance Tested: n. OK None Checklist for HP reduction type DRS Functional check of regulator Leak check Functional check of analysis gas heating (if any) Functional check of HP reduction heating (if any) Checklist for cylinder rack Check and documentation of cylinder pressures of carrier gas/gases and calibration gas Functional test of cylinder heating...
  • Page 32 Table section 3 / annual routine maintenance Tested: n. OK None Checklist for HP reduction type DRS Functional check of regulator Leak check Functional check of analysis gas heating (if any) Functional check of HP reduction heating (if any) Checklist for cylinder rack Check and documentation of cylinder pressures of carrier gas/gases and calibration gas Functional test of cylinder heating...
  • Page 33 Table section 3 / annual routine maintenance Tested: n. OK None Checklist for HP reduction type DRS Functional check of regulator Leak check Functional check of analysis gas heating (if any) Functional check of HP reduction heating (if any) Checklist for cylinder rack Check and documentation of cylinder pressures of carrier gas/gases and calibration gas Functional test of cylinder heating...
  • Page 34: Test Gas Analyses

    Table section 4 Table section 4 Test gas analyses To be performed at: • Metrological check • (Re)calibration Entries made by: • Trained personnel • RMG Service PGC9300 Maintenance Logbook · EN07 · 2018, June 26th...
  • Page 35 Table section 4 / test gas analyses Procedure for metrological check / (re)calibration 1. Performance of basic calibration with logging of the - retention times (RTZ) - response factors (RFZ / GLK) 2. Analysis of the test gases 3. Verification of the chromatograms / data transmission Current outputs only need to be checked if they are used for official transmission of fiscal metering readings.
  • Page 36 Tabellenteil section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
  • Page 37 Table section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
  • Page 38 Tabellenteil section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
  • Page 39 Table section 4 / test gas analyses Verification of the chromatogram of the internal calibration gas Verification of the chromatogram of test gas 1 Verification of the chromatogram of test gas 2 Verification of the chromatogram of test gas 3 Verification of the chromatogram of test gas 4 Verification of the chromatogram of test gas 5 Verification of the chromatogram of test gas 6...
  • Page 40 Tabellenteil section 4 / test gas analyses Procedure for metrological check / (re)calibration 1. Performance of basic calibration with logging of the - retention times (RTZ) - response factors (RFZ / GLK) 2. Analysis of the test gases 3. Verification of the chromatograms / data transmission Current outputs only need to be checked if they are used for official transmission of fiscal metering readings.
  • Page 41 Table section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
  • Page 42 Tabellenteil section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
  • Page 43 Table section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
  • Page 44 Tabellenteil section 4 / test gas analyses Verification of the chromatogram of the internal calibration gas Verification of the chromatogram of test gas 1 Verification of the chromatogram of test gas 2 Verification of the chromatogram of test gas 3 Verification of the chromatogram of test gas 4 Verification of the chromatogram of test gas 5 Verification of the chromatogram of test gas 6...
  • Page 45 Table section 4 / test gas analyses Procedure for metrological check / (re)calibration 1. Performance of basic calibration with logging of the - retention times (RTZ) - response factors (RFZ / GLK) 2. Analysis of the test gases 3. Verification of the chromatograms / data transmission Current outputs only need to be checked if they are used for official transmission of fiscal metering readings.
  • Page 46 Tabellenteil section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
  • Page 47 Table section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
  • Page 48 Tabellenteil section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
  • Page 49 Table section 4 / test gas analyses Verification of the chromatogram of the internal calibration gas Verification of the chromatogram of test gas 1 Verification of the chromatogram of test gas 2 Verification of the chromatogram of test gas 3 Verification of the chromatogram of test gas 4 Verification of the chromatogram of test gas 5 Verification of the chromatogram of test gas 6...
  • Page 50 Tabellenteil section 4 / test gas analyses Procedure for metrological check / (re)calibration 1. Performance of basic calibration with logging of the - retention times (RTZ) - response factors (RFZ / GLK) 2. Analysis of the test gases 3. Verification of the chromatograms / data transmission Current outputs only need to be checked if they are used for official transmission of fiscal metering readings.
  • Page 51 Table section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
  • Page 52 Tabellenteil section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
  • Page 53 Table section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
  • Page 54 Tabellenteil section 4 / test gas analyses Verification of the chromatogram of the internal calibration gas Verification of the chromatogram of test gas 1 Verification of the chromatogram of test gas 2 Verification of the chromatogram of test gas 3 Verification of the chromatogram of test gas 4 Verification of the chromatogram of test gas 5 Verification of the chromatogram of test gas 6...
  • Page 55 Table section 4 / test gas analyses Procedure for metrological check / (re)calibration 1. Performance of basic calibration with logging of the - retention times (RTZ) - response factors (RFZ / GLK) 2. Analysis of the test gases 3. Verification of the chromatograms / data transmission Current outputs only need to be checked if they are used for official transmission of fiscal metering readings.
  • Page 56 Tabellenteil section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
  • Page 57 Table section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
  • Page 58 Tabellenteil section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
  • Page 59 Table section 4 / test gas analyses Verification of the chromatogram of the internal calibration gas Verification of the chromatogram of test gas 1 Verification of the chromatogram of test gas 2 Verification of the chromatogram of test gas 3 Verification of the chromatogram of test gas 4 Verification of the chromatogram of test gas 5 Verification of the chromatogram of test gas 6...
  • Page 60 Tabellenteil section 4 / test gas analyses Procedure for metrological check / (re)calibration 1. Performance of basic calibration with logging of the - retention times (RTZ) - response factors (RFZ / GLK) 2. Analysis of the test gases 3. Verification of the chromatograms / data transmission Current outputs only need to be checked if they are used for official transmission of fiscal metering readings.
  • Page 61 Table section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
  • Page 62 Tabellenteil section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
  • Page 63 Table section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
  • Page 64 Tabellenteil section 4 / test gas analyses Verification of the chromatogram of the internal calibration gas Verification of the chromatogram of test gas 1 Verification of the chromatogram of test gas 2 Verification of the chromatogram of test gas 3 Verification of the chromatogram of test gas 4 Verification of the chromatogram of test gas 5 Verification of the chromatogram of test gas 6...
  • Page 65 Table section 4 / test gas analyses Procedure for metrological check / (re)calibration 1. Performance of basic calibration with logging of the - retention times (RTZ) - response factors (RFZ / GLK) 2. Analysis of the test gases 3. Verification of the chromatograms / data transmission Current outputs only need to be checked if they are used for official transmission of fiscal metering readings.
  • Page 66 Tabellenteil section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
  • Page 67 Table section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
  • Page 68 Tabellenteil section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
  • Page 69 Table section 4 / test gas analyses Verification of the chromatogram of the internal calibration gas Verification of the chromatogram of test gas 1 Verification of the chromatogram of test gas 2 Verification of the chromatogram of test gas 3 Verification of the chromatogram of test gas 4 Verification of the chromatogram of test gas 5 Verification of the chromatogram of test gas 6...
  • Page 70 Tabellenteil section 4 / test gas analyses Procedure for metrological check / (re)calibration 1. Performance of basic calibration with logging of the - retention times (RTZ) - response factors (RFZ / GLK) 2. Analysis of the test gases 3. Verification of the chromatograms / data transmission Current outputs only need to be checked if they are used for official transmission of fiscal metering readings.
  • Page 71 Table section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
  • Page 72 Tabellenteil section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
  • Page 73 Table section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
  • Page 74 Tabellenteil section 4 / test gas analyses Verification of the chromatogram of the internal calibration gas Verification of the chromatogram of test gas 1 Verification of the chromatogram of test gas 2 Verification of the chromatogram of test gas 3 Verification of the chromatogram of test gas 4 Verification of the chromatogram of test gas 5 Verification of the chromatogram of test gas 6...
  • Page 75 Table section 4 / test gas analyses Procedure for metrological check / (re)calibration 1. Performance of basic calibration with logging of the - retention times (RTZ) - response factors (RFZ / GLK) 2. Analysis of the test gases 3. Verification of the chromatograms / data transmission Current outputs only need to be checked if they are used for official transmission of fiscal metering readings.
  • Page 76 Tabellenteil section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
  • Page 77 Table section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
  • Page 78 Tabellenteil section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
  • Page 79 Table section 4 / test gas analyses Verification of the chromatogram of the internal calibration gas Verification of the chromatogram of test gas 1 Verification of the chromatogram of test gas 2 Verification of the chromatogram of test gas 3 Verification of the chromatogram of test gas 4 Verification of the chromatogram of test gas 5 Verification of the chromatogram of test gas 6...
  • Page 80 Tabellenteil section 4 / test gas analyses Procedure for metrological check / (re)calibration 1. Performance of basic calibration with logging of the - retention times (RTZ) - response factors (RFZ / GLK) 2. Analysis of the test gases 3. Verification of the chromatograms / data transmission Current outputs only need to be checked if they are used for official transmission of fiscal metering readings.
  • Page 81 Table section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
  • Page 82 Tabellenteil section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
  • Page 83 Table section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
  • Page 84 Tabellenteil section 4 / test gas analyses Verification of the chromatogram of the internal calibration gas Verification of the chromatogram of test gas 1 Verification of the chromatogram of test gas 2 Verification of the chromatogram of test gas 3 Verification of the chromatogram of test gas 4 Verification of the chromatogram of test gas 5 Verification of the chromatogram of test gas 6...
  • Page 85 Subject to technical modification For further information please visit our website: www.rmg.com or contact your local sales support office to learn more about the RMG products. RMG Messtechnik GmbH Otto-Hahn-Straße 5 35510 Butzbach, Germany Phone: +49 (0) 6033 897 – 0 Fax: +49 (0) 6033 897 –...