Process Gas Chromatograph PGC 9300 MAINTENANCE LOGBOOK Reliable Measurement of Gas Status: 2018, JUNE 26TH Version:...
Page 2
Contact Manufacturer Our customer service is available for technical queries Address RMG Messtechnik GmbH Otto-Hahn-Straße 5 D-35510 Butzbach +49 6033 897 – 0 Telefon switchboard +49 6033 897 – 0 Telefon service +49 6033 897 – 173 Telefon spare parts +49 6033 897 –...
Page 3
Your PGC9300 PGC 930 Maintenace Logbook PGC manufacturer number PGC year of manufacture Meter serial number GC 9300 manufacturer number GC 9300 year of manufacture GC 9300 software version Calorific value Gas quality Approved for Operator Station Commissioning Type / carrier Column temperature Column pressure...
The setpoints for the operating parameters are entered on the first page of this book by the test laboratory during commissioning and shall remain unchanged during operation, unless altered by an RMG service technician. The actual values of the operating parameters can be checked at any time on the GC 9300 analyser under the Status->Meter tab or using RMGViewGC.
1. Instructions for Keeping the Maintenance Logbook 1.1.2. Cylinder pressures The pressure in the gas cylinders is checked directly at the high-pressure gauges of the relevant pressure reducing unit at regular intervals, in other words at every maintenance check, and the information is then documented in the relevant table in Table Section 1 of this logbook.
The annual inspection of the PGC 930x must be carried out on the basis of the checklist (Table Section 4) by persons who have received appropriate training from RMG to provide such maintenance. 2.2. Metrological check / (re)calibration Once the maintenance activities described in Section 2 are complete, the device will subsequently undergo routine recalibration.
Table section 1 Table Section 1 Cylinder pressures To be performed at: • Every inspection • Every maintenance Entries made by: • Operator • RMG Service PGC9300 Maintenance Logbook · EN07 · 2018, June 26th...
Page 9
Table section 1 / cylinder pressures : Cylinder changed : Device faulty Carriergas I Int. calibration gas Carrier gas II Date Initials Pressure / bar Pressure / bar Pressure / bar PGC9300 Maintenance Logbook · EN07 · 2018, June 26th...
Page 10
Table section 1 / cylinder pressures : Cylinder changed : Device faulty Carriergas I Int. calibration gas Carrier gas II Date Initials Pressure / bar Pressure / bar Pressure / bar PGC9300 Maintenance Logbook · EN07 · 2018, June 26th...
Page 11
Table section 1 / cylinder pressures : Cylinder changed : Device faulty Carriergas I Int. calibration gas Carrier gas II Date Initials Pressure / bar Pressure / bar Pressure / bar PGC9300 Maintenance Logbook · EN07 · 2018, June 26th...
Table section 2 Maintenance measures, general To be performed at: • Annual routine maintenance • Software update • Metrological check • (Re)calibration • Repair Entries made by: • Trained personnel • RMG Service PGC9300 Maintenance Logbook · EN07 · 2018, June 26th...
Page 13
Table section 2 / maintenance measures, general Maintenance measure Performed by Performed on Measures performed Routine maintenance Table 3 Software update Metrological check Table 4 (Re)calibration Table 4 Defect See below Description of problem Measure Description of problem Measure Signature PGC9300 Maintenance Logbook ·...
Page 14
Table section 2 / maintenance measures, general Maintenance measure Performed by Performed on Measures performed Routine maintenance Table 3 Software update Metrological check Table 4 (Re)calibration Table 4 Defect See below Description of problem Measure Description of problem Measure Signature PGC9300 Maintenance Logbook ·...
Page 15
Table section 2 / maintenance measures, general Maintenance measure Performed by Performed on Measures performed Routine maintenance Table 3 Software update Metrological check Table 4 (Re)calibration Table 4 Defect See below Description of problem Measure Description of problem Measure Signature PGC9300 Maintenance Logbook ·...
Page 16
Table section 2 / maintenance measures, general Maintenance measure Performed by Performed on Measures performed Routine maintenance Table 3 Software update Metrological check Table 4 (Re)calibration Table 4 Defect See below Description of problem Measure Description of problem Measure Signature PGC9300 Maintenance Logbook ·...
Page 17
Table section 2 / maintenance measures, general Maintenance measure Performed by Performed on Measures performed Routine maintenance Table 3 Software update Metrological check Table 4 (Re)calibration Table 4 Defect See below Description of problem Measure Description of problem Measure Signature PGC9300 Maintenance Logbook ·...
Page 18
Table section 2 / maintenance measures, general Maintenance measure Performed by Performed on Measures performed Routine maintenance Table 3 Software update Metrological check Table 4 (Re)calibration Table 4 Defect See below Description of problem Measure Description of problem Measure Signature PGC9300 Maintenance Logbook ·...
Page 19
Table section 2 / maintenance measures, general Maintenance measure Performed by Performed on Measures performed Routine maintenance Table 3 Software update Metrological check Table 4 (Re)calibration Table 4 Defect See below Description of problem Measure Description of problem Measure Signature PGC9300 Maintenance Logbook ·...
Page 20
Table section 2 / maintenance measures, general Maintenance measure Performed by Performed on Measures performed Routine maintenance Table 3 Software update Metrological check Table 4 (Re)calibration Table 4 Defect See below Description of problem Measure Description of problem Measure Signature PGC9300 Maintenance Logbook ·...
Page 21
Table section 2 / maintenance measures, general Maintenance measure Performed by Performed on Measures performed Routine maintenance Table 3 Software update Metrological check Table 4 (Re)calibration Table 4 Defect See below Description of problem Measure Description of problem Measure Signature PGC9300 Maintenance Logbook ·...
Page 22
Table section 2 / maintenance measures, general Maintenance measure Performed by Performed on Measures performed Routine maintenance Table 3 Software update Metrological check Table 4 (Re)calibration Table 4 Defect See below Description of problem Measure Description of problem Measure Signature PGC9300 Maintenance Logbook ·...
Table section 3 Table section 3 Annual routine maintenance To be performed at: • Annual routine maintenance Entries made by: • Trained personnel • RMG Service PGC9300 Maintenance Logbook · EN07 · 2018, June 26th...
Page 24
Table section 3 / annual routine maintenance Tested: n. OK None Checklist for HP reduction type DRS Functional check of regulator Leak check Functional check of analysis gas heating (if any) Functional check of HP reduction heating (if any) Checklist for cylinder rack Check and documentation of cylinder pressures of carrier gas/gases and calibration gas Functional test of cylinder heating...
Page 25
Table section 3 / annual routine maintenance Tested: n. OK None Checklist for HP reduction type DRS Functional check of regulator Leak check Functional check of analysis gas heating (if any) Functional check of HP reduction heating (if any) Checklist for cylinder rack Check and documentation of cylinder pressures of carrier gas/gases and calibration gas Functional test of cylinder heating...
Page 26
Table section 3 / annual routine maintenance Tested: n. OK None Checklist for HP reduction type DRS Functional check of regulator Leak check Functional check of analysis gas heating (if any) Functional check of HP reduction heating (if any) Checklist for cylinder rack Check and documentation of cylinder pressures of carrier gas/gases and calibration gas Functional test of cylinder heating...
Page 27
Table section 3 / annual routine maintenance Tested: n. OK None Checklist for HP reduction type DRS Functional check of regulator Leak check Functional check of analysis gas heating (if any) Functional check of HP reduction heating (if any) Checklist for cylinder rack Check and documentation of cylinder pressures of carrier gas/gases and calibration gas Functional test of cylinder heating...
Page 28
Table section 3 / annual routine maintenance Tested: n. OK None Checklist for HP reduction type DRS Functional check of regulator Leak check Functional check of analysis gas heating (if any) Functional check of HP reduction heating (if any) Checklist for cylinder rack Check and documentation of cylinder pressures of carrier gas/gases and calibration gas Functional test of cylinder heating...
Page 29
Table section 3 / annual routine maintenance Tested: n. OK None Checklist for HP reduction type DRS Functional check of regulator Leak check Functional check of analysis gas heating (if any) Functional check of HP reduction heating (if any) Checklist for cylinder rack Check and documentation of cylinder pressures of carrier gas/gases and calibration gas Functional test of cylinder heating...
Page 30
Table section 3 / annual routine maintenance Tested: n. OK None Checklist for HP reduction type DRS Functional check of regulator Leak check Functional check of analysis gas heating (if any) Functional check of HP reduction heating (if any) Checklist for cylinder rack Check and documentation of cylinder pressures of carrier gas/gases and calibration gas Functional test of cylinder heating...
Page 31
Table section 3 / annual routine maintenance Tested: n. OK None Checklist for HP reduction type DRS Functional check of regulator Leak check Functional check of analysis gas heating (if any) Functional check of HP reduction heating (if any) Checklist for cylinder rack Check and documentation of cylinder pressures of carrier gas/gases and calibration gas Functional test of cylinder heating...
Page 32
Table section 3 / annual routine maintenance Tested: n. OK None Checklist for HP reduction type DRS Functional check of regulator Leak check Functional check of analysis gas heating (if any) Functional check of HP reduction heating (if any) Checklist for cylinder rack Check and documentation of cylinder pressures of carrier gas/gases and calibration gas Functional test of cylinder heating...
Page 33
Table section 3 / annual routine maintenance Tested: n. OK None Checklist for HP reduction type DRS Functional check of regulator Leak check Functional check of analysis gas heating (if any) Functional check of HP reduction heating (if any) Checklist for cylinder rack Check and documentation of cylinder pressures of carrier gas/gases and calibration gas Functional test of cylinder heating...
Table section 4 Table section 4 Test gas analyses To be performed at: • Metrological check • (Re)calibration Entries made by: • Trained personnel • RMG Service PGC9300 Maintenance Logbook · EN07 · 2018, June 26th...
Page 35
Table section 4 / test gas analyses Procedure for metrological check / (re)calibration 1. Performance of basic calibration with logging of the - retention times (RTZ) - response factors (RFZ / GLK) 2. Analysis of the test gases 3. Verification of the chromatograms / data transmission Current outputs only need to be checked if they are used for official transmission of fiscal metering readings.
Page 36
Tabellenteil section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
Page 37
Table section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
Page 38
Tabellenteil section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
Page 39
Table section 4 / test gas analyses Verification of the chromatogram of the internal calibration gas Verification of the chromatogram of test gas 1 Verification of the chromatogram of test gas 2 Verification of the chromatogram of test gas 3 Verification of the chromatogram of test gas 4 Verification of the chromatogram of test gas 5 Verification of the chromatogram of test gas 6...
Page 40
Tabellenteil section 4 / test gas analyses Procedure for metrological check / (re)calibration 1. Performance of basic calibration with logging of the - retention times (RTZ) - response factors (RFZ / GLK) 2. Analysis of the test gases 3. Verification of the chromatograms / data transmission Current outputs only need to be checked if they are used for official transmission of fiscal metering readings.
Page 41
Table section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
Page 42
Tabellenteil section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
Page 43
Table section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
Page 44
Tabellenteil section 4 / test gas analyses Verification of the chromatogram of the internal calibration gas Verification of the chromatogram of test gas 1 Verification of the chromatogram of test gas 2 Verification of the chromatogram of test gas 3 Verification of the chromatogram of test gas 4 Verification of the chromatogram of test gas 5 Verification of the chromatogram of test gas 6...
Page 45
Table section 4 / test gas analyses Procedure for metrological check / (re)calibration 1. Performance of basic calibration with logging of the - retention times (RTZ) - response factors (RFZ / GLK) 2. Analysis of the test gases 3. Verification of the chromatograms / data transmission Current outputs only need to be checked if they are used for official transmission of fiscal metering readings.
Page 46
Tabellenteil section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
Page 47
Table section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
Page 48
Tabellenteil section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
Page 49
Table section 4 / test gas analyses Verification of the chromatogram of the internal calibration gas Verification of the chromatogram of test gas 1 Verification of the chromatogram of test gas 2 Verification of the chromatogram of test gas 3 Verification of the chromatogram of test gas 4 Verification of the chromatogram of test gas 5 Verification of the chromatogram of test gas 6...
Page 50
Tabellenteil section 4 / test gas analyses Procedure for metrological check / (re)calibration 1. Performance of basic calibration with logging of the - retention times (RTZ) - response factors (RFZ / GLK) 2. Analysis of the test gases 3. Verification of the chromatograms / data transmission Current outputs only need to be checked if they are used for official transmission of fiscal metering readings.
Page 51
Table section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
Page 52
Tabellenteil section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
Page 53
Table section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
Page 54
Tabellenteil section 4 / test gas analyses Verification of the chromatogram of the internal calibration gas Verification of the chromatogram of test gas 1 Verification of the chromatogram of test gas 2 Verification of the chromatogram of test gas 3 Verification of the chromatogram of test gas 4 Verification of the chromatogram of test gas 5 Verification of the chromatogram of test gas 6...
Page 55
Table section 4 / test gas analyses Procedure for metrological check / (re)calibration 1. Performance of basic calibration with logging of the - retention times (RTZ) - response factors (RFZ / GLK) 2. Analysis of the test gases 3. Verification of the chromatograms / data transmission Current outputs only need to be checked if they are used for official transmission of fiscal metering readings.
Page 56
Tabellenteil section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
Page 57
Table section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
Page 58
Tabellenteil section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
Page 59
Table section 4 / test gas analyses Verification of the chromatogram of the internal calibration gas Verification of the chromatogram of test gas 1 Verification of the chromatogram of test gas 2 Verification of the chromatogram of test gas 3 Verification of the chromatogram of test gas 4 Verification of the chromatogram of test gas 5 Verification of the chromatogram of test gas 6...
Page 60
Tabellenteil section 4 / test gas analyses Procedure for metrological check / (re)calibration 1. Performance of basic calibration with logging of the - retention times (RTZ) - response factors (RFZ / GLK) 2. Analysis of the test gases 3. Verification of the chromatograms / data transmission Current outputs only need to be checked if they are used for official transmission of fiscal metering readings.
Page 61
Table section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
Page 62
Tabellenteil section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
Page 63
Table section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
Page 64
Tabellenteil section 4 / test gas analyses Verification of the chromatogram of the internal calibration gas Verification of the chromatogram of test gas 1 Verification of the chromatogram of test gas 2 Verification of the chromatogram of test gas 3 Verification of the chromatogram of test gas 4 Verification of the chromatogram of test gas 5 Verification of the chromatogram of test gas 6...
Page 65
Table section 4 / test gas analyses Procedure for metrological check / (re)calibration 1. Performance of basic calibration with logging of the - retention times (RTZ) - response factors (RFZ / GLK) 2. Analysis of the test gases 3. Verification of the chromatograms / data transmission Current outputs only need to be checked if they are used for official transmission of fiscal metering readings.
Page 66
Tabellenteil section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
Page 67
Table section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
Page 68
Tabellenteil section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
Page 69
Table section 4 / test gas analyses Verification of the chromatogram of the internal calibration gas Verification of the chromatogram of test gas 1 Verification of the chromatogram of test gas 2 Verification of the chromatogram of test gas 3 Verification of the chromatogram of test gas 4 Verification of the chromatogram of test gas 5 Verification of the chromatogram of test gas 6...
Page 70
Tabellenteil section 4 / test gas analyses Procedure for metrological check / (re)calibration 1. Performance of basic calibration with logging of the - retention times (RTZ) - response factors (RFZ / GLK) 2. Analysis of the test gases 3. Verification of the chromatograms / data transmission Current outputs only need to be checked if they are used for official transmission of fiscal metering readings.
Page 71
Table section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
Page 72
Tabellenteil section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
Page 73
Table section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
Page 74
Tabellenteil section 4 / test gas analyses Verification of the chromatogram of the internal calibration gas Verification of the chromatogram of test gas 1 Verification of the chromatogram of test gas 2 Verification of the chromatogram of test gas 3 Verification of the chromatogram of test gas 4 Verification of the chromatogram of test gas 5 Verification of the chromatogram of test gas 6...
Page 75
Table section 4 / test gas analyses Procedure for metrological check / (re)calibration 1. Performance of basic calibration with logging of the - retention times (RTZ) - response factors (RFZ / GLK) 2. Analysis of the test gases 3. Verification of the chromatograms / data transmission Current outputs only need to be checked if they are used for official transmission of fiscal metering readings.
Page 76
Tabellenteil section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
Page 77
Table section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
Page 78
Tabellenteil section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
Page 79
Table section 4 / test gas analyses Verification of the chromatogram of the internal calibration gas Verification of the chromatogram of test gas 1 Verification of the chromatogram of test gas 2 Verification of the chromatogram of test gas 3 Verification of the chromatogram of test gas 4 Verification of the chromatogram of test gas 5 Verification of the chromatogram of test gas 6...
Page 80
Tabellenteil section 4 / test gas analyses Procedure for metrological check / (re)calibration 1. Performance of basic calibration with logging of the - retention times (RTZ) - response factors (RFZ / GLK) 2. Analysis of the test gases 3. Verification of the chromatograms / data transmission Current outputs only need to be checked if they are used for official transmission of fiscal metering readings.
Page 81
Table section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
Page 82
Tabellenteil section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
Page 83
Table section 4 / test gas analyses Relevant for Test gas 1 Test gas 2 calibration? Type Error Type Error ↓ actual absolute actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N-butane Neopentane Isopentane N-pentane Hexane (C6+) Heptane Octane Nonane...
Page 84
Tabellenteil section 4 / test gas analyses Verification of the chromatogram of the internal calibration gas Verification of the chromatogram of test gas 1 Verification of the chromatogram of test gas 2 Verification of the chromatogram of test gas 3 Verification of the chromatogram of test gas 4 Verification of the chromatogram of test gas 5 Verification of the chromatogram of test gas 6...
Page 85
Subject to technical modification For further information please visit our website: www.rmg.com or contact your local sales support office to learn more about the RMG products. RMG Messtechnik GmbH Otto-Hahn-Straße 5 35510 Butzbach, Germany Phone: +49 (0) 6033 897 – 0 Fax: +49 (0) 6033 897 –...