1.2.9

Disposal of packaging material

Dispose of the material in an environmentally friendly manner in accordance with
national standards and directives.
1.2.10

Storage

Avoid extended periods of storage. The PGC 9300 is a high-precision measuring
device that should not be stored for an extended time. A new factory calibration
may be necessary after an extended period of storage (longer than 4 weeks). In-
spect the gas chromatograph PGC 9300 for damage and correct function after
storage. Contact the RMG service department to arrange for inspection of the de-
vice after a storage period of longer than one year. For this purpose, send the de-
vice to RMG.
However, if storage is necessary, the following must be observed:
• The PGC 9300 must be positioned upright after it has been unloaded
• The PGC 9300 must be secured against movement and tipping over
• Storage in an air-conditioned and dry environment is required due to the
sensitivity to moisture
• All supply and discharge lines for gas must be sealed during storage (condi-
tion as delivered). If applicable, seal the lines with sealing plugs or dummy
plugs.
• Installation must be carried out by qualified personnel only.
• If storage times of longer than 4 weeks are necessary, a permanent carrier
gas flush must be provided.
An alternative storage is described in chapter 0