Gas Flow Distribution - RMG PGC 9300 Operating Manual

Measuring element
Hide thumbs Also See for PGC 9300:
Table of Contents

Advertisement

3 Gas chromatograph PGC 9300

3.4 Gas flow distribution

The purpose of the gas distribution is to connect one of the four gas streams to the
analyzer and to supply the carrier gas.
In order to avoid contamination of the selected gas stream due to potential leak-
36
age of the valves, a so-called "Double Block and Bleed" arrangement of valves
was selected.
Figure 7: Gas (distribution principle diagram) shows a corresponding arrangement
in which Gas Stream 2 is selected. It shows that the volumes between the valves
of gas streams that are not selected are vented to the atmosphere. As a result,
any occurring leaks cannot contaminate the analysis gas stream.
1
2
3
1. sample gas
2. calibration gas
3. test gas
4. safe shut off valve
Manual PGC 9300 EN07 June, 18th 2018
7
5
5
5
5. gas filter
1. Betriebsgas
6. sample gas outlet
2. Kalibriergas
7. off gas
3. Referenzgas
4. Magnetventil
Figure 7: Gas (distribution principle diagram)
4
5. Filter
6. Analysengas
7. Entlüftung
6

Advertisement

Table of Contents
loading

Table of Contents