- Application of pressure (Figure 11: Application of pressure)
The sample solenoid valve opens the micro valve for the gas sample with carrier
gas and applies a control pressure of 5.5 bar. The micro solenoid valve is blocked
and a defined sample volume is enclosed in the sample loop. The switching sole-
noid valve regulates the pressure in the sample loop with carrier gas.
CARRIER GAS
5.5bar
PS
PRESSURE
REGULATOR
SAMPLE VENT
CARRIER GAS 5.5bar
CARRIER GAS REGULATED
SAMPLE
INJECTION
SAMPLE
CARRIER GAS
SWITCH
SAMPLE LOOP
Figure 11: Application of pressure
3 Gas chromatograph PGC 9300
TCD
REFERENCE COLUMN
SAMPLE COLUMN
INJECTION
SAMPLE
Manual PGC 9300 EN07 June, 18th 2018
41
SAMPLE IN