1 About this manual
Gas Clean Moisture Filter (Agilent)
Note
A color indicator shows whether the filter must be replaced. Fields with
corresponding colors for ORIGINAL and SATURATED are printed on the
23
filter cap.
The filter must be changed when the display field begins to change color.
If the indicator completely changes color to SATURATED, damage to the
measuring element is possible. The service department must be contacted
in this case.
Caution
We urgently recommend contacting RMG customer service to replace the
filter on the PGC 9300 in order to minimize the risk of damage to the meas-
uring element.