Appendix 2: Error messages

Appendix 2: Error messages
Errors are outputted by the analytical computer as a text message with a specific
number. The error numbers are outputted in the print logs.
68
A complete list is provided in Appendix E of the GC 9300 manual.
Only the error messages directly related to the analysis should be addressed here.

Appendix 2 A: Error during operation

Number
Text
14 - 17
Current output - #
limit
50
Measuring element
timeout
120
Ana: Retention time
121
Ana: unnormalized
total
122
Ana: Concentration
123,
Ana: Ho min/max
82,
Ana: CO
Ana: Wo min/max
124,
Ana: Rho,n min/max
127
130
Measuring gas P-min
/ max
131
Carrier gas pressure -
I
Manual PGC 9300 EN07 June, 18th 2018
Description
Current < 0/2 mA and/or > 21mA
The measuring element no longer pro-
vides valid data.
Communication error
The measuring element works cor-
rectly, but there are no measurements
received by the analytical computer.
The retention time of one or multiple
gas components deviates impermissi-
bly from the base values.
(Retention times in the "Detail" screen
for the individual components, permissi-
ble deviation under "Computing param-
eters / limit values ana.,cal.")
With standardization to 100%, the limit
value is exceeded ("Calibration results /
area totals" and "Calibration parame-
ters/limit values cal.").
Working range of the modules ex-
ceeded and/or undercut
Measuring range exceeded and/or un-
min/max
dercut
2
Depending on the inlet that is switched
on, the inlet pressure of the measur-
ing/calibrating or reference gas is out-
side of the tolerance.
The carrier gas inlet pressure is outside
of the tolerance.
Cause
- Check of limit values in the "De-
tail" screen under "inputs and
outputs"
- Communication error
- Error in the measuring element
- Wire break
- Incorrectly specified TCP/IP ad-
dress in the analytical computer
- Faulty pressure/temperature val-
ues (with error 50)
- Impermissible gas composition
(with error 121, 122)
- Defective module
- Impermissible gas composition
- Pressure/temperature error (50)
- Defective module
- impermissible gas composition
- Check of limit values in the "De-
tail" screen under "Components
and parameters"
- Incorrect setting
- Cylinder pressure
- Check "Status" screen
- Incorrect setting
- Cylinder pressure
- Check "Status" screen