Chapter 7 RDO/DO Calibration
the log and calibration data can be exported from the meter to a printer or computer.
Air Dissolved Oxygen Calibration
calibration can be performed in a BOD bottle.
Note: In most calibration screens, press f1 (esc) to return to the measurement mode without saving the calibration.
to highlight DO/RDO – Channel and press f2 (select).
2. Press
or 102.3 % if using a polarographic DO probe.
Thermo Scientific Orion Star A329 Portable pH/ISE/Conductivity/RDO/DO Meter Reference Guide
f1 (cal). In dual and multi channel measurement displays, press
or
to highlight Air and press f3 (select).
Reading is stable. Accept Auto Calibration Value and 101.7 % if using an RDO optical DO probe
f3 (cal done) to export the data to the calibration log or press f2 (print) to export the data to the calibration
f3 (start).
or