Selecting The Program Language; Configuring The Flow Sensor - GE Sievers 900 Series Operation And Maintenance Manual

On-line total organic carbon analyzers
Table of Contents

Advertisement

7.
Press Back when you are done.
If the Analyzer is not powered on when a TOC Autozero is programmed to occur, the
Note:
scheduled TOC Autozero is skipped. If the composition of the water is changing too
rapidly at the time the TOC Autozero is scheduled, the TOC Autozero will be stopped.
In both cases, the next TOC Autozero will occur at its regularly programmed time.

Selecting the Program Language

If you have loaded a second program language into the Analyzer, you can switch between the two
languages by following these steps:
1.
Select the Maintenance tab.
2.
Press the Advanced button.
3.
Press the Advanced Setup button.
4.
Press the Select Language button. Two buttons will display; the choices depend on which
languages you have loaded into the Analyzer. Select the desired language to continue.
Note
If you change the program language by mistake and need help navigating the menus to
switch back to your native language, see "Problems Changing the Program Language" on
page 212.

Configuring the Flow Sensor

The Analyzer automatically detects the presence of the sample stream when it is working in On-
Line mode. When the water flow is interrupted, the Analyzer issues an error and stops analysis. To
change the Flow Sensor setting, follow these steps:
1.
Select the Maintenance tab.
2.
Press the Advanced button.
3.
Press the Advanced Setup button.
4.
Press the Flow Sensor button.
GE Analytical Instruments ©2009
118 of 236
DLM 90488-01 Rev. A