GE Sievers 900 Series Operation And Maintenance Manual page 168

On-line total organic carbon analyzers
Table of Contents

Advertisement

Autoreagent Calibration Worksheet
Company Name ___________________________
Analyst Name ____________________________
Analyzer Serial Number ____________________
Standard
25 ppm TOC
50 ppm TOC
25 ppm IC Standard
IC
TC
Autoreagent calibration pass/fail based upon the following criteria:
The median peak time of the 50 ppm TOC Standard is from 20 to 100 seconds less than the
median peak time of the 25 ppm TOC Standard.
The RSDs of the valid expected IC and TC concentration measurements are 3% or less.
The RSDs of the IC and TC preliminary measurements are 5% or less
Autoreagent Calibration Results:
Autoreagent Calibration Action:
GE Analytical Instruments ©2009
Expected Conc.
RSD (%)
(ppm)
Passed
Applied
168 of 236
Date ________________________
Firmware Version _____________
Peak Time (sec)
Prelim. Meas.
(ppm)
Failed
Canceled
RSD (%)
DLM 90488-01 Rev. A