Autoreagent Calibration Worksheet
Company Name ___________________________
Analyst Name ____________________________
Analyzer Serial Number ____________________
Standard
25 ppm TOC
50 ppm TOC
25 ppm IC Standard
IC
TC
Autoreagent calibration pass/fail based upon the following criteria:
•
The median peak time of the 50 ppm TOC Standard is from 20 to 100 seconds less than the
median peak time of the 25 ppm TOC Standard.
•
The RSDs of the valid expected IC and TC concentration measurements are 3% or less.
•
The RSDs of the IC and TC preliminary measurements are 5% or less
Autoreagent Calibration Results:
Autoreagent Calibration Action:
GE Analytical Instruments ©2009
Expected Conc.
RSD (%)
(ppm)
Passed
Applied
168 of 236
Date ________________________
Firmware Version _____________
Peak Time (sec)
Prelim. Meas.
(ppm)
Failed
Canceled
RSD (%)
DLM 90488-01 Rev. A