If you know the specific flow rates you want to use for each reagent, you can manually set the flow
rates, according to Table 5 on page 98 and Table 6 on page 99.
Note:
Analysis must be stopped prior to changing the reagent flow rates.

Using the Autoreagent Function

Set up the Autoreagent function by following these steps:
1.
Select the Setup tab.
2.
Press the Configure Mode button (see Figure 13).
3.
Make sure the Reagent button is set to Auto. The Analyzer will automatically calculate the
flow rates for both reagents.
4.
To modify the automatically calculated acid flow rate, press the Acid button and select an
adjustment value: +10%, +20%, -10%, or -20%. You also can return the value to Auto.
5.
To modify the automatically calculated oxidizer flow rate, press the Oxidizer button and
select an adjustment value: +10%, +20%, -10%, or -20%. You also can return the value to
Auto.
Configure
GE Analytical Instruments ©2009
Reagent
Manual
Acid
Oxidizer
0.05
0.02
Reps
Rejects
5
1
Menu
Figure 13: Selecting the Reagent Flow Method
4 Mar 2004 08:16
Auto
Manual
97 of 236