Establishing A New Administrator Account For Dataguard - GE Sievers 900 Series Operation And Maintenance Manual

On-line total organic carbon analyzers
Table of Contents

Advertisement

Once DataGuard is enabled, you will immediately be required to log in to the Analyzer with the
default Administrator User ID and Password:
User ID: ADMIN
Password: GEAI
Proceed to the section "Establishing a New Administrator Account for DataGuard" below to
complete the DataGuard activation.
After installation is complete, you can add unique user accounts to the Analyzer. Refer to Chapter 5,
"Password Protection and DataGuard" for more information.

Establishing a New Administrator Account for DataGuard

After you log in to the Analyzer for the first time after activating DataGuard, create a new
administrator-level account and inactivate the default administrator account to ensure Analyzer
security. (Refer to Chapter 5, "Password Protection and DataGuard" for complete DataGuard
information.)
1.
Press the Menu button and select the DataGuard tab.
2.
Press the Add User button.
3.
Specify a new User ID for the administrator and press the Enter button.
4.
Specify the Password for the administrator User ID and press Enter.
5.
Verify the Password and press Enter.
6.
Press the User Level button and select Administrator.
7.
Press the Password Expired button and select False.
8.
Press the Back button.
9.
Press the Logout button. You will now use the new administrator account and delete the
default account.
10. Press the Login button and enter the new User ID and Password you just created.
11. Press the Menu button and select the DataGuard tab.
12. Press the Edit User button.
13. Use the arrow buttons to highlight the default administrator account and press OK.
14. Press the User Status button and select Inactive.
GE Analytical Instruments ©2009
81 of 236
DLM 90488-01 Rev. A

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents