Setting Up Analyzer Input And Output; Setting Up Analog Output - GE Sievers 900 Series Operation And Maintenance Manual

On-line total organic carbon analyzers
Table of Contents

Advertisement

5.
Indicate the destination port by pressing the USB or Serial button to start the data export.
The data is exported in comma-separated format (.csv) and contains the following fields: Analysis
mode, Date, Time, TOC, TC, IC, Acid, Oxidizer, Reagent mode, IC cell temperature, TC cell
temperature, and ICR state. Most spreadsheet programs, such as Microsoft Excel, can open the
.csv file without having to go through an import process.

Setting Up Analyzer Input and Output

Menu selections under the I/O tab allow you to configure data for alarms, analog output, binary
input, serial output and printing. Note that older 900 Portable Analyzers may not have a 4-20 mA
output.
I/O
Main

Setting Up Analog Output

Select the output range for the 4-20 mA analog output by following the steps below. Instructions for
wiring the 4-20mA output can be found in the "Installation" chapter.
1.
Select the I/O tab.
GE Analytical Instruments ©2009
Setup
Data
I/O
Serial
Printer
Binary Input
Analog Out
Start Analysis
Figure 17: The I/O Tab
4 Mar 2004 08:23
Maintenance
DataGuard
Alarm 1
Alarm 2
105 of 236