Table 12 Standards Required For Single-Point Calibration; Table 13 Standards Required For Calibration Verification; Table 14 Standards Required For System Suitability Verification - GE Sievers 900 Series Operation And Maintenance Manual

On-line total organic carbon analyzers
Table of Contents

Advertisement

Table 12: Standards Required for Single-Point Calibration
Calibration Blank
1, 5, 10, 25, or 50 ppm TOC (as KHP)
1, 5, 10, 25, or 50 ppm IC (as Na
*Note: 1ppm = 1 mg C/L, 1ppb = 1 µg C/L
Calibration verification standards also should be purchased in a concentration that is appropriate
for your application; sets include one vial of reagent water blank and one vial each of TOC and IC in
the selected concentration. The available concentrations are shown in Table 13.
Table 13: Standards Required for Calibration Verification
Verification Blank
500 ppb or 1, 2, 5, 10, 25, or 50 ppm TOC (as sucrose)
500 ppb or 1, 2, 5, 10, 25, or 50 ppm IC (as Na
*Note: 1ppm = 1 mg C/L, 1ppb = 1 µg C/L
System suitability standards sets include the standards shown in Table 14.
Table 14: Standards Required for System Suitability Verification
Reagent Water Blank - Rw
500 ppb TOC (as sucrose) - Rs
500 ppb TOC (as benzoquinone) - Rss
*Note: 1ppm = 1 mg C/L, 1ppb = 1 µg C/L
GE Analytical Instruments ©2009
TOC Calibration Standards
IC Calibration Standards
CO
2
Verification Standards (TOC and IC)
138 of 236
)
3
CO
)
2
3
DLM 90488-01 Rev. A