To avoid exposure to the chemical reagents, wear acid-resistant gloves, protective
clothing, and safety goggles or a face shield when changing the reagent supplies.
Reagent containers are for single-use only — do not refill. Refilling or reusing
reagent containers will void all Analyzer and parts warranties and nullify any
performance claims.
1.
Open the Analyzer front panel by turning the two latches and swinging the panel door to
the left.
2.
Locate the reagent enclosure, located at the top left-hand side of the Analyzer as you face
the Analyzer. Loosen the set screw and remove the reagent enclosure cover.
3.
Remove the reagent containers from the packaging. Record the expiration date of the acid
and oxidizer and the volume of the oxidizer in the Notes column on the Maintenance
Worksheet (see Table 18 on page 193); you will need to enter these dates into the
Analyzer in Step 8.
4.
Locate the acid supply line, indicated by the Acid label. Attach the PEEK nut to the acid
reagent container and tighten finger-tight.
5.
Slide the valve on the acid container to the open position by pushing the green button all
the way in. Refer to the labeling on the container for proper positioning.
6.
Slide the acid container into the reagent enclosure, with the valve pointing down and
toward the front of the reagent enclosure.
7.
Locate the oxidizer supply line, indicated by the Oxidizer label. Attach the PEEK nut to the
oxidizer container and tighten finger-tight.
8.
Slide the valve on the oxidizer container to the open position by pushing the green button
all the way in. Refer to the labeling on the container for proper positioning.
9.
Slide the oxidizer container into the reagent enclosure, with the valve pointing down and
toward the front of the reagent enclosure.
10. Replace the reagent enclosure cover and tighten the set screw until it is finger-tight.
GE Analytical Instruments ©2009