GE Sievers 900 Series Operation And Maintenance Manual page 200

On-line total organic carbon analyzers
Table of Contents

Advertisement

flash memory drives, follow the Sievers 900 TOC Analyzers link on our Web page,
www.GEInstruments.com.
50 - Unstable Measurements Error
Message: "Unstable measurements detected during calibration/verification. Please refer to the 900
Oper. and Maint. Manual for recommendations."
Explanation: This error displays when the results from calibration/verification are inappropriate and
calibration/verification cannot continue. Follow the suggestions in the section called "Step 2: Visual
Inspection" on page 201 before continuing with analysis. You may need to contact Technical
Support to confirm that the Analyzer is functioning properly.
51 - History Flash Error
Message: "A problem has been detected in the history data flash. Please contact GE."
Explanation: The Analyzer experienced a problem when trying to store data in the flash memory. It
is possible a section of the flash memory is damaged. To protect against losing data, stop analysis
and contact Technical Support.
52 - User List Error
Message: "The user list is full. Archive the user list before adding more users."
Explanation: DataGuard allows a maximum of 100 users. This error displays when 100 user
accounts have been created. If you have inactive accounts in the user list, you can archive users
(DataGuard tab
Archive Users) to remove them from the list; you will then be able to add
additional users.
53 - History Data Error
Message: "The history data is full. Archive the history data before taking more measurements."
Explanation: This error displays when the data history is completely filled. Export (Data
tab
Export) or Print (Data tab
Print the data) before proceeding.
54 - Audit Trail Error
Message: "The audit trail is full. Archive the audit trail before taking more measurements."
Explanation: This error displays when the audit trail is completely filled. To comply with 21 CFR Part
11, you must either Export (DataGuard tab
Export AuditTrail) or Print (DataGuard tab
Print
Audit Trail) the audit trail before proceeding.
55 - ICR Error
Message: "A problem has been detected with the ICR. Check the ICR indicator lights. Refer to the
900 ICR Manual for recommendations."
Explanation: This error generally displays when the ICR is not creating a vacuum but is set to Inline.
Refer to the 900 ICR Operation and Maintenance Manual for troubleshooting suggestions.
56 - Acid Syringe Error
Message: "Error detected with the acid syringe. Inspect the syringe and cables. Refer to the 900
Oper. and Maint. Manual for recommendations."
GE Analytical Instruments ©2009
200 of 236
DLM 90488-01 Rev. A