Figure 23 Selecting The Concentration For A Single-Point Calibration - GE Sievers 900 Series Operation And Maintenance Manual

On-line total organic carbon analyzers
Table of Contents

Advertisement

When you are ready to calibrate the Analyzer, follow these steps:
1.
If the Analyzer is taking measurements, press the Stop Analysis button.
2.
If you have an ICR unit, place the lever into the Bypass position. If you leave the lever in the
Inline position, a warning message will display later.
3.
Select the Maintenance tab and press the Calibrate button.
4.
Press the Single Pt Cal button.
5.
If you have not performed a reagent flush in the last 8 hours, you must perform one now.
Make sure the Analyzer is connected to a water supply and click Yes.
If you have performed a reagent flush in the last 8 hours, click No to skip the flush and
proceed to Step 6.
6.
When the flush is complete, click OK to continue. Remove the protective cap from the
standards vials, and avoid touching the septa to protect against introducing foreign
particles.
7.
Press the Cal. Standard button, select the concentration of the standard you will be
measuring, and push Next.
At this point, the prompts on the Analyzer's display screen will help guide you through the
calibration process.
Calibrate
Figure 23: Selecting the Concentration for a Single-Point Calibration
8.
Insert the Calibration Blank into the IOS System, and press Next.
GE Analytical Instruments ©2009
Cal. Standard
25.0 PPM
Menu
17 Feb 2004 08:09
1.0 PPM
5.0 PPM
10.0 PPM
25.0 PPM
50.0 PPM
144 of 236
DLM 90488-01 Rev. A