Additional Installation Equipment; Step 2: Complete The Identification Records - GE Sievers 900 Series Operation And Maintenance Manual

On-line total organic carbon analyzers
Table of Contents

Advertisement

• Rubber plug for IOS waste line (only needed for 900 On-Line or 900 Portable Analyzers)
5.
Reagent package containing:
• Acid reagent cartridge
• Oxidizer reagent cartridge
Note:
Do not start the Analyzer until the reagents have been installed and the DI water
reservoir has been filled with DI water.
6.
Certificate of Calibration
7.
DataGuard Activation key (optional)

Additional Installation Equipment

To install the Sievers 900 On-Line TOC Analyzer, the following equipment also is required:
Mounting hardware to support the Analyzer (see "Step 3: Select a Location for the
Analyzer" on page 62)
Insulated wire (22-12 American Wire Gauge, rated to 300 Volts) for analog and alarm
outputs (optional; see "Installing the Analog Outputs and Alarms" on page 68)
Analog (4-20 mA) recorder (optional)
Insulated wire (18-12 American Wire Gauge, rated to 300 Volts) for AC power conduit (see
"Connecting to a Power Supply" on page 75)
An external circuit breaker or switch that disconnects both poles of the supply voltage,
rated appropriately (optional)
Grounding strap for ESD protection

Step 2: Complete the Identification Records

Complete the Identification Records section on "Identification Records" on page 2 by recording the
date of installation and the Analyzer serial number (found on the left side of the Analyzer).
GE Analytical Instruments ©2009
61 of 236
DLM 90488-01 Rev. A

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents