Table 9: Audit Trail Output Format - GE Sievers 900 Series Operation And Maintenance Manual

On-line total organic carbon analyzers
Table of Contents

Advertisement

Field
Field #
Contents
1
User ID
2
Entry ID
Number
3
Date and
Time
4
Action
5
Old Value
6
New Value
Logging Out
When users have finished their activities on the Analyzer, logging out prevents a second user from
accessing the system under the first user's account. This practice helps preserve the integrity of
the audit trail. Logging out also minimizes the risk of unauthorized access by other personnel.
To log out, press the Logout button on any display screen. If the Logout button is not displayed on
the current screen, press the Back or Menu button until any of the main tabs (Setup, Data, I/O,
Maintenance, or DataGuard) display, and then press the Logout button.
GE Analytical Instruments ©2009

Table 9: Audit Trail Output Format

Format/Notes
11 alpha-numeric characters (columns 1-11)
9 numeric characters (columns 13-21)
dd MMM yyyy hh:mm:ss (columns 23-33)
20 alpha-numeric characters (columns 44-63)
Alpha-numeric characters (columns 65-end of
line)
Alpha-numeric characters (columns 65-end of
line)
133 of 236
Example
ADMIN 01
693090237
17 OCT 2004
PASSWORD SET
QUALITY 01
QUALITY 01
DLM 90488-01 Rev. A