Sample Flow Rate Calibration Worksheet
Company Name ___________________________
Analyst Name ____________________________
Analyzer Serial Number ____________________
Sample pump tubing replaced before sample flow rate calibration
Enter data for the sample flow rate calibration:
Expected
mL/min:
0.50
Flow rate calibration pass/fail based upon the following criteria:
The Adjusted sample flow rate is 0.48 to 0.52 mL/min
Sample Flow Rate Calibration Results:
Sample Flow Rate Calibration Action:
GE Analytical Instruments ©2009
Measured
________
Measured Flow Rate - 0.50
% Diff =
---------------------------------------------------------------------- -
0.50
Passed
Applied
Date ________________________
Firmware Version _____________
(Y/N): ______
% Diff
________
×
100%
Failed
Canceled