GE Sievers 900 Series Operation And Maintenance Manual page 164

On-line total organic carbon analyzers
Table of Contents

Advertisement

Verification Worksheet
Company Name ___________________________
Analyst Name ____________________________
Analyzer Serial Number ____________________
Concentration of Standard: _______
Reagent Flush Performed (Y/N): ______
Enter data for the TOC Verification:
Expected
Blank
N/A
________
________
Enter data for the IC Verification:
Expected
________
________
Measured Concentration - Expected Standard Concentration
% Diff =
-----------------------------------------------------------------------------------------------------------------------------------------------------------------
Pass criteria:
RSD for each vial is 3% or less.
TOC % Diff is ±7% or less for a 500 ppb or 1 ppm TOC standard.
TOC % Diff is ±5% or less for a 2, 5, 10, 25, or 50 ppm TOC standard.
IC % Diff is ±7 or less% for a 500 ppb or 1 ppm IC standard.
IC % Diff is ±5% or less for a 2, 5, 10, 25, or 50 ppm IC standard.
Verification Results:
GE Analytical Instruments ©2009
(Line 1)
(Line 2)
Measured
________
________
Measured
________
Expected Standard Concentration
Passed
Failed
Date ________________________
Firmware Version _____________
RSD
________
________
RSD
________
164 of 236
% Diff
________
________
% Diff
________
×
100%
DLM 90488-01 Rev. A