2.5.2 The "STOP" Key

With the "stop" key (10), (see Figure 17), the run can be interrupted at any time. After pressing the
key, the centrifuge decelerates, with the respective pre-selected intensity, down to a standstill.
Figure 17

2.6 Imbalance Detection

If the rotor is not equally loaded, the drive will turn off, during acceleration. The rotor decelerates to
a standstill.
When in the display, "time" (A-3), the word "error" (M11) along with the number "01" (Figure 18) may
appear, the weight difference of the samples is too large. Weigh out the samples, precisely.
Load the rotor as described in: Chapters 2.1.2 and 2.1.3.
When in the display, "time" (A-3), the word "error" (M11) along with the number "02" (see Figure 18)
may appear, for the following reason:
*The imbalance sensor is defective.