Warranty; Installation Of The Centrifuge; Unpacking The Centrifuge; Space Requirements - Hermle Z 206 A Operating Manual

Hide thumbs Also See for Z 206 A:
Table of Contents

Advertisement

PRODUCT DESCRIPTION

1.6.6 Warranty

The centrifuge has been subjected to thorough testing and quality control. In the unlikely case of any
manufacturing faults occurring, the centrifuge and rotors are covered by warranty, for a period of two
years, from date of delivery. This warranty becomes invalid in any case of mishandling, damage
and/or negligence and further in any case of usage of inappropriate spare parts and / or accessories
or unauthorized modification of the unit.
Technical modification rights are reserved, by the manufacturer, in regards to
technical improvement!

1.7 Installation of the Centrifuge

1.7.1 Unpacking the Centrifuge

Model Z 206 A is supplied in a carton.
Remove the strap retainer, open the carton, and remove the centrifuge. The instruction manual must
be kept with the centrifuge, at all times!

1.7.2 Space Requirements

The centrifuge should be installed on an even, solid surface, if possible on a laboratory cabinet /
table or some other solid vibration free surface.
During centrifugation, the centrifuge must be placed in a way, that there is a minimum space of
30 cm/11.81in on each side of the unit, according to EN 61010-2-020 standards.
Do not place the centrifuge next to a window or a heater where it could be disposed to excessive
heat, as the performance of the unit is based on an ambient temperature of 23°C/73.4°F.

1.7.3 Installation

Follow these steps:
• Check whether the power supply corresponds with the one specified on the manufacturer's rating
label, mounted on the rear panel.
• The line voltage circuit breaker is max. 10 A (Type K), slowly release for commonly used
instruments.
• In case of emergency, there must be an emergency switch off installed outside of the room, in order
to disconnect the power supply from the unit.
• Connect the centrifuge, with the mains.
(The socket for the power cord must be easy to reach, respectively easy to disconnect).
-Switch on, by using the mains power switch (2.2).
Open the lid, by using the button LID.
• Remove the transport securing device of the motor.
6
Z206A_V1.18_eng
© Hermle Labortechnik GmbH

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents