Check the integrity of the seals of the aerosol-tight rotor lids or caps, before each use.
-
Use only aerosol-tight rotor lids or caps, if the seals are undamaged and clean.
-
Replace the seals of aerosol-tight lids and caps, after 5 autoclaving cycles.
-
Never store aerosol-tight rotors or buckets closed.

3.1.2 Cleaning and Disinfecting of the Unit

1. Open the lid, before turning off the unit. Disconnect from the power supply.
2. Open the rotor nut, by turning the rotor key counter-clockwise.
3. Remove the rotor.
4. For cleaning and disinfecting of the unit and the rotor chamber, use the cleaner previously
described, (see Chapter 3.1.1).
5. Clean all accessible areas of the device and its accessories, including the power cord, with a
damp cloth.
6. Wash the rubber seals and rotor chamber thoroughly, with water.
7. Rub the dry rubber seals with glycerol or talc, to prevent these from becoming brittle. Other
components of the unit, e.g. the lid lock, motor shaft and rotor, should not be greased.
8. Dry the motor shaft with a soft, dry and lint-free cloth.
9. Examine the unit and accessories for damage.
Remove adherent dust, at least every 6 months, from the ventilation slots in the centrifuge, by using
a soft brush.
*Before doing so, please switch off the unit and disconnect from the power supply.

3.1.3 Cleaning and Disinfecting of the Rotor

1. Clean and disinfect the rotors, rotor lids and adapters with the cleaner mentioned in: (Chapter
3.1.1).
2. Use a bottle brush to clean and disinfect the rotor bores.
3. Rinse the rotors, rotor lids and adapters with clear water. Particularly, the drillings of the angle
rotors.
4. When drying the rotors and accessories, set on a towel. Place the angle rotors, with bores
down, to dry.
5. Dry the rotor cone with a soft, dry and lint-free cloth, check for damage. Do not grease the rotor
cone.

3.1.4 Disinfection of Aluminium - Rotors

In case of infectious material spilling into the centrifuge, the rotor and rotor chamber have to be
disinfected, promptly after the run. Rotors may be autoclaved, at a maximum temperature of
121°C/250°F.
.

3.1.5 Disinfection of PP-Rotors

The recommended time for autoclaving: 15 – 20 min at 121°C/250°F, (1 bar)
ATTENTION: The sterilization time of 20 min. must not be exceeded. Continuous sterilization
will cause reduction in the mechanical resistance, of the plastic material.
Before autoclaving the PP-rotor and adapter, thoroughly clean to avoid the burning of dirty residues.