PRODUCT DESCRIPTION

1.8 Basic Adjustment

1.8.1 Adjustment of the Rotor Type

Before the first operation resp., after each rotor change, choose the rotor type needed.
Find the rotor type on the printed order number, on the rotor.
Example:
Angle Rotor:
220.96 V01 ≙ rotortype "96"
Swing-Out Rotor: 220.68 V04 ≙ rotortype "68"
Switch on the unit and open the lid. Press simultaneously, the keys "lid" (7) and "stop" (10). In the
display, "rpm/rcf" (A-1) appears the old previous rotor type, i. e. "96". With the potentiometer (1),
choose the used rotor now, (see Figure 1). To select the data in the unit, please press the "start" (9)
key.
*Now all important rotor parameters are stored in the centrifuge.
Figure 1
Attention: Set rotor must always match with the used rotor, otherwise the unit can be
damaged.
The rotor type can be checked during the run by pressing the key "lid" (7).