For any resulting damage, the company, Hermle Labortechnik, is not liable. 1.3 Brief Description The unit type Z 287 A is a non-refrigerated universal centrifuge, which we offer in two voltage variations 230V or 120V. The centrifuge can be used with swing-out rotors and angle rotors.
Installation of the Centrifuge 1.5.1 Unpacking the Centrifuge Model Z 287 A is supplied in a carton. Remove the strap retainer, open the carton and remove the centrifuge. The instruction manual must be kept with the centrifuge at all times! 1.5.2 Space Requirements...
PRODUCT DESCRIPTION 1.6 Signs- and Indication of the Centrifuge 1.6.1 Product Nameplate (Example) Company Address: Hermle Labortechnik GmbH, Siemensstr. 25, D-78564 Wehingen TYPE: Type Designation of the Product REF: Order No. of the Product Serial No. of the Product Manufacturer Date of Manufacture MAX.
EN 61010-2-020. • The HERMLE Z 287 A is non "explosion-proof" and must not be operated in explosion- endangered areas or locations. Centrifugation of flammable, explosive, radioactive, or such substances, which chemically react with high energy, is strictly prohibited. If used in such environment, this is at the users own expense.
If In case of any questions, please contact the manufacturer! 2.1.4 Rotor Identification The model Z 287 A comes with an automatic rotor identification system. Place the rotor on the motor shaft as described in 2.1.1 and close the centrifuge lid. The centrifuge immediately recognizes the rotor and shows its item number on the display.
Health risk from contaminated equipment, rotors and accessories In case of returning the centrifuge for repairing to the manufacturer, please know the following: The centrifuge must be decontaminated and cleaned, before shipment to Hermle Labortechnik, for the protection of persons, environment and material.
6.4 RoHS II Compliance HERMLE Labortechnik GmbH, Siemensstraße 25, 78564 Wehingen, hereby declares and certifies that all components manufactured are RoHS II compliant, according to the definition and restrictions given by the European Parliament Directive 2011/65/EC. This restricts the use of certain hazardous substances in electrical and electronic equipmen Z287A_V1.19_...
Typ Designation Z 206 A; Z 207 A; Z 207 H, Z 167 M; Z 207 M, Z 216 M; Z 287 A; Z 306; Z 326; Z 366; Z 446;Z 216 MK; Z 32 HK; Z 326 K; Z 366 K; Z 36 HK; Z 446 K; ZK 496 Einschlägige EG-Richtlinien / Normen...
Page 35
APPENDIX Table 1: Technical Data Manufacturer HERMLE Labortechnik GmbH, 78564 Wehingen Z 287 A Type Dimensions 35 cm Width 43 cm Depth 26 cm Height 18 kg Weight without rotor 14,000 rpm Max. Speed 6 x 50 ml Max. Volume 16,058 x g Max.