The "Stop" Key; Imbalance Detection - Hermle Z 207 H User Manual

Table of Contents

Advertisement

OPERATING
Figure 19

2.7.2 The "STOP" KEY

With the "stop" key (10), the run time can be interrupted, at any time, (see Figure 20). After pressing
the key, the centrifuge decelerates with the respective pre-selected intensity, down to a standstill.
Figure 20

2.8 Imbalance Detection

In case the rotor is not equally loaded, the drive will turn off, during acceleration. The rotor decelerates
to a standstill.
When in the display "time" (A-3), the word "error" (M11) along with the number "01" appears, the
weight difference of the samples are too large. Weigh out the samples exactly!
Load the rotor as described in Chapter: 2.1.2
When inside the display "time" (A-3), the word "error" along with the number "02" appears, (see Figure
21). Potential reason for this error can be: The imbalance switch is defective.
20
Z207H_V1.22_eng
© Hermle Labortechnik GmbH

Advertisement

Table of Contents
loading

Table of Contents