Hermle Z 206 A Operating Manual
Hide thumbs Also See for Z 206 A:
Table of Contents

Advertisement

Operating Manual for
Small Centrifuge
Z 206 A
© Hermle Labortechnik GmbH
Z206A_V1.18_eng

Advertisement

Table of Contents
loading

Summary of Contents for Hermle Z 206 A

  • Page 1 Operating Manual for Small Centrifuge Z 206 A © Hermle Labortechnik GmbH Z206A_V1.18_eng...
  • Page 3: Table Of Contents

    Pre-Selection of Running Time ........................15 2.4.3 Pre-Selection of Brake Intensity and Acceleration..................15 Starting and Stopping the Centrifuge ......................16 2.5.1 Starting the Centrifuge........................... 16 2.5.2 The "STOP" Key ............................17 Imbalance Detection ............................17 3. MAINTENANCE ......................... 18 © Hermle Labortechnik GmbH Z206A_V1.18_eng...
  • Page 4 Table 2: Permissible Net Weight ..........................V Table 3: Max. Speed and RCF-Values for Permissible Rotors ................V Table 4: Acceleration and Deceleration Times ......................V Table 5: Error Messages ............................VI Table 6: Symbol / Abbreviations..........................VI Notes ............................I Z206A_V1.18_eng © Hermle Labortechnik GmbH...
  • Page 5: Product Description

    For any resulting damage, the company, Hermle Labortechnik, is not liable. 1.3 Brief Description The unit type Z 206 A is a non refrigerated universal centrifuge, which we offer in two voltage variations 230V or 120V. The centrifuge can be used with swing-out and angle rotors.
  • Page 6: Operating And Display Elements

    Centrifugation Time Lid Release start Start Centrifugation stop Stop Centrifugation 1.5.1 LCD-Display The following picture shows the individual elements of the LCD-Display. Display Fields: Display Field – "rpm/rcf" Display Field – "acc/dec" Display Field– "time" Z206A_V1.18_eng © Hermle Labortechnik GmbH...
  • Page 7: Signs And Indications Of The Centrifuge

    Instructions for Disposal (see Chapter 6, S. 23) Direction of Rotation – clockwise rotation for the rotor drive Reference for Loading Rotors 1.6.2 Product Nameplate (Example) Company Address: Hermle Labortechnik GmbH, Siemensstr. 25, D-78564 Wehingen TYPE: Type Designation of the Product REF: Order No. of the Product...
  • Page 8: Warning And Information Signs

    Take off mains plug before opening the housing or the emergency release Power Input This device may only be operated by a trained professional. Carefully, read the operating manual and be familiar with the functions of the device. Z206A_V1.18_eng © Hermle Labortechnik GmbH...
  • Page 9: Declaration Of Atex (94/9/Eg)

    PRODUCT DESCRIPTION 1.6.4 Declaration of ATEX (94/9/EG) The HERMLE Z 206 A is not "explosion-proof" and must therefore not be operated in explosion- endangered areas or locations. During centrifugation, it is prohibited to stay within the safety zone of 30 cm/12in around the centrifuge or deposit hazardous substances, within this area. Centrifugation...
  • Page 10: Warranty

    1.7 Installation of the Centrifuge 1.7.1 Unpacking the Centrifuge Model Z 206 A is supplied in a carton. Remove the strap retainer, open the carton, and remove the centrifuge. The instruction manual must be kept with the centrifuge, at all times! 1.7.2 Space Requirements...
  • Page 11: Basic Adjustment

    *Now all important rotor parameters are stored in the centrifuge. Figure 1 Attention: Set rotor must always match with the used rotor, otherwise the unit can be damaged. The rotor type can be checked during the run by pressing the key "lid" (7). © Hermle Labortechnik GmbH Z206A_V1.18_eng...
  • Page 12: Access To The Mode: "Operating Data

    Now release both keys. As a result, a display test is administered for approx. 5 seconds. All possible indications will appear at the same time (see Figure 2). Figure 2 Z206A_V1.18_eng © Hermle Labortechnik GmbH...
  • Page 13: Retrieving Operating Data

    (see figure 3). The first two numbers indicate the appeared errors ongoing from 00 to 99, the last two numbers indicate the error code. Please look up in "table 5: error messages" (See APPENDIX VI). Switch off the centrifuge, to change back to the normal program mode. Figure 3 © Hermle Labortechnik GmbH Z206A_V1.18_eng...
  • Page 14: Operation

    Do not operate the centrifuge with rotors or buckets, that show any signs of corrosion or mechanical damage. Do not operate with extremely corrosive substances, that could damage the rotor, buckets and materials. In case of any questions, please contact the manufacturer! Z206A_V1.18_eng © Hermle Labortechnik GmbH...
  • Page 15: Loading Angle Rotors

    Swing Out Rotors can be removed during operation, only if all locations are filled in with either 4 buckets or 4 carriers – do not mix up buckets and carriers!! Figure 9: WRONG Figure 10: RIGHT © Hermle Labortechnik GmbH Z206A_V1.18_eng...
  • Page 16: Loading And Overloading Of Rotors

    The power switch is located on the left-hand side of the ground plate, on the unit, (see Figure 11). Figure 11: Power Switch ATTENTION: After turning on the power switch, open the centrifuge lid first, before starting the centrifuge. Z206A_V1.18_eng © Hermle Labortechnik GmbH...
  • Page 17: Lid

    (see Figure 12). ATTENTION Before closing the lid, please check if the rotor is tightened, and all 6 buckets are inside the swing out rotor. © Hermle Labortechnik GmbH Z206A_V1.18_eng...
  • Page 18: Pre-Selection

    The pre-selection, of the RCF-value, is the same value. The setting range is between 20 xg and the maximum permissible centrifugal force, of the rotor. The maximum speed of the Z 206 A is 6,000 rpm resp. 4,427 xg. Please refer to: "Table 3: Max. Speed and RCF", (See APPENDIX V). This table shows the approved rotors with their max.
  • Page 19: Pre-Selection Of Running Time

    9 is equivalent to the shortest possible brake time and the value 0 to longest possible brake time. See Table 5: "Acceleration and Deceleration Times", (See APPENDIX V). This table shows the acceleration and deceleration times, for the acceleration and deceleration stages 0 to 9, for each rotor. © Hermle Labortechnik GmbH Z206A_V1.18_eng...
  • Page 20: Starting And Stopping The Centrifuge

    When the respective pre-selected running time has ended, the centrifuge will stop automatically or the run can be interrupted, with the “stop” (10) key. For all number marked text, please refer to Figure 16. Figure 16 Z206A_V1.18_eng © Hermle Labortechnik GmbH...
  • Page 21: The "Stop" Key

    Load the rotor as described in: Chapters 2.1.2 and 2.1.3. When in the display, “time” (A-3), the word “error” (M11) along with the number “02“ (see Figure 18) may appear, for the following reason: *The imbalance sensor is defective. Figure 18 © Hermle Labortechnik GmbH Z206A_V1.18_eng...
  • Page 22: Maintenance

    "TÜV Nord CERT GmbH, Certification Body Consumer Products, Essen (Germany)", in accordance with Annex AA of IEC 61010-2-020. The certificates can be downloaded on our webpage, www.hermle-labortechnik.de. Aerosol-tight rotors and buckets are marked with the label, "aerosol- tight". ATTENTION: Autoclaving, mechanical stresses and contamination, by chemicals or other aggressive solvents, can impair the aerosol-tightness of the rotors and buckets.
  • Page 23: Cleaning And Disinfecting Of The Unit

    ATTENTION: The sterilization time of 20 min. must not be exceeded. Continuous sterilization will cause reduction in the mechanical resistance, of the plastic material. Before autoclaving the PP-rotor and adapter, thoroughly clean to avoid the burning of dirty residues. © Hermle Labortechnik GmbH Z206A_V1.18_eng...
  • Page 24: Glass Breakage

    Transparent rotor lids and caps, made of PC or PP, as well as rotors, tube racks and adapters of PP, have a maximum operating time of up to 3 years, from first time use. Conditions For the Operating Time: Proper use, damage-free condition, recommended care. Z206A_V1.18_eng © Hermle Labortechnik GmbH...
  • Page 25: Trouble Shouting

    • Pull the string slightly and the lid will open. ATTENTION: Don’t put your hands in the rotor chamber, while it is still spinning! • Push the plastic stopper back into the unit, to continue normal function. Figure 19 © Hermle Labortechnik GmbH Z206A_V1.18_eng...
  • Page 26: Description Of The Error Message System

    In case of a return of a centrifuge, we also ask you to send its power cord. This eliminates the risk of a faulty power cord. If no power cord is attached to the centrifuge, a new one will be delivered and charged. We ask for your understanding. Z206A_V1.18_eng © Hermle Labortechnik GmbH...
  • Page 27: Transport, Storage, Disposal

    RoHS II Compliance HERMLE Labortechnik GmbH, Siemensstraße 25, 78564 Wehingen, hereby declares and certifies that all components manufactured are RoHS II compliant, according to the definition and restrictions given by the European Parliament Directive 2011/65/EC. This restricts the use of certain hazardous substances in electrical and electronic equipment.
  • Page 28: Appendix

    Table 3: Max. Speed and RCF-Values for Permissible Rotors ......... V Table 4: Acceleration and Deceleration Times ..............V Table 5: Error Messages ....................VI Table 6: Symbol / Abbreviations ..................VI Redemption Form / Decontamination Certificate ............VIII Z206A_V1.18_eng © Hermle Labortechnik GmbH...
  • Page 29: Ec Conformity Declaration

    Typ Designation Z 206 A; Z 207 A; Z 207 H; Z 207 M; Z 216 M; Z 306; Z 326; Z 366; Z 446; Z 216 MK; Z 32 HK; Z 326 K; Z 366 K; Z 36 HK; Z 446 K Einschlägige EG-Richtlinien / Normen...
  • Page 30 Table 1: Technical Data HERMLE Labortechnik GmbH, 78564 Wehingen Manufacturer Z 206 A Type Dimensions 28 cm Width 37 cm Depth 26 cm Height 10,5 kg Weight without rotor Max. Speed 6000 min Max. Volume 6 x 50 ml 4427 x g Max.
  • Page 31 Level 9 Level 0 Level 9 220.68 V04 220.95 V06 221.54 V01 221.55 V01 220.96 V01 220.97 V01 221.37 V01 in seconds Acceleration Time Deceleration Time from 0 min -> n from n -> 0 min © Hermle Labortechnik GmbH Z206A_V1.18_eng...
  • Page 32 Open lid while the rotor is running Over-speed Under-voltage in the intermediate circuit Sticking relay Table 6: Symbol / Abbreviations Symbol / Abbreviation Unit Description Revolutions Per Minute n (=rpm) [min Relative Centrifugal Force RZB(=rcf) [x g] Polypropylene Polycarbonate Z206A_V1.18_eng © Hermle Labortechnik GmbH...
  • Page 33 © Hermle Labortechnik GmbH Z206A_V1.18_eng...
  • Page 34 Have these substances reached the equipment/assembly? …………………………..  Yes  No If so, which ones: ________________________________________________________________ Description of the measures for the decontamination of the listed parts: I confirm the proper decontamination: Company/Dept ._____________ Place and Date: ______________________________ Signature of the authorized person: ________________________________________ VIII Z206A_V1.18_eng © Hermle Labortechnik GmbH...
  • Page 35 © Hermle Labortechnik GmbH Z206A_V1.18_eng...
  • Page 36 Z206A_V1.18_eng © Hermle Labortechnik GmbH...
  • Page 37: Notes

    Notes © Hermle Labortechnik GmbH Z206A_V1.18_eng...
  • Page 39 © Hermle Labortechnik GmbH Z206A_V1.18_eng...
  • Page 40 HERMLE Labortechnik GmbH Siemensstraße 25 78564 Wehingen Tel: 0 74 26-96 22-17 Fax: 0 74 26-96 22-49 Email: vertrieb@hermleLT.de Internet: http://www.hermle-labortechnik.de Technical rights reserved ©HERMLE Labortechnik GmbH 2018...