4. TROUBLE SHOUTING

4.1 Error Messages: Problem / Solution

The error messages are listed to help localize possible errors faster.
The possible error referred to in this chapter may not always be the case, as they are only theoretically
occurring errors and solutions.
Always keep us informed about any kind of error occurring, which is not listed in this chapter. With this
information provided, we are able to improve and complete this operation manual.
Many thanks in advance for your support.
HERMLE Labortechnik GmbH

4.2 Survey of Possible Error Messages and Solutions

4.2.1 Lid Release during Power Failure (Emergency Lid Release)

In case of power failure or malfunction, the lid of the centrifuge can be opened manually, in order to
protect samples.
Please proceed as follows:
• Switch the centrifuge off and unplug the power cord.
• At the left-hand side of the centrifuge housing, there is a plastic stopper, (see Figure 19).
Remove the stopper, fastened to it, is a string which is connected to the electronic lid lock.
• Pull the string slightly and the lid will open.
ATTENTION: Don't put your hands in the rotor chamber, while it is still spinning!
• Push the plastic stopper back into the unit, to continue normal function.
Figure 19
© Hermle Labortechnik GmbH
Z206A_V1.18_eng