For any resulting damage, the company, Hermle Labortechnik, is not liable. 1.3 Brief Description The unit type Z 207 A is a non refrigerated universal centrifuge, which can be used with 100 – 230V country-specific power cords. The centrifuge can only be used with one angle rotor.
1.6 Signs- and Indication of the Centrifuge 1.6.1 General Direction of Rotation – clockwise rotation for the rotor drive 1.6.2 Product Nameplate (Example) Company Address: Hermle Labortechnik GmbH, Siemensstr. 25, D-78564 Wehingen TYPE: Type Designation of the Product REF: Article Number of the Product...
EN 61010-2-020. • The HERMLE Z 207 A is non "explosion-proof" and must not be operated in explosion- endangered areas or locations. Centrifugation of flammable, explosive, radioactive, or such substances, which chemically react with high energy, is strictly prohibited. If used in such environment, this is at the users own expense.
Installation of the Centrifuge 1.7.1 Unpacking the Centrifuge Model Z 207 A is supplied in a carton. Remove the strap retainer, open the carton and remove the centrifuge. The instruction manual must be kept with the centrifuge at all times! 1.7.2 Space Requirements...
Open the centrifuge lid and shut off the main switch. Now switch on again the main switch.For approximately 3 seconds “Z 207 A“ will be shown in the display. Press during this time the keys “time“ (6) and “lid“ (10) simultaneously. As a result, a display test is administered for approx. 3 seconds.
It is the same with the pre-selection of the RCF-Value. The setting range is between 20 xg and the maximum permissible centrifugal force of the rotor. The permissible max. speed of the Z 207 A with the rotor 221.57 V01 is 6,800 rpm or 4,445 xg. ATTENTION: Please notice the maximum permissible revolutions of the test tubes (Producer Indication)! See also chapter 3.1.5...
Health risk from contaminated equipment, rotors and accessories In case of returning the centrifuge for repairing to the manufacturer, please know the following: The centrifuge must be decontaminated and cleaned, before shipment to Hermle Labortechnik, for the protection of persons, environment and material.
RoHS II Compliance HERMLE Labortechnik GmbH, Siemensstraße 25, 78564 Wehingen, hereby declares and certifies that all components manufactured are RoHS II compliant, according to the definition and restrictions given by the European Parliament Directive 2011/65/EC. This restricts the use of certain hazardous substances in electrical and electronic equipmen Z207A_V1.17...
Typenbezeichnung Typ Designation Z 206 A; Z 207 A; Z 216 M; Z 306; Z 326; Z 366; Z 446; Z 216 MK; Z 32 HK; Z 326 K; Z 366 K; Z 36 HK; Z 446 K Einschlägige EG-Richtlinien / Normen Relevant EC Directives / Standards 98/79/EG (Anhang/Annex III);...
Page 35
APPENDIX Table 1: Technical Data HERMLE Labortechnik GmbH, 78564 Wehingen Manufacturer Z 207 A Type Dimensions 28 cm Width 35 cm Depth 24 cm Height 8,0 kg Weight without rotor Max. Speed 6,800 min 8 x 15 ml Max. Volume 4,445 x g Max.