PRODUCT DESCRIPTION

1.8 Basic Adjustment

1.8.1 Adjustment of the Rotor Type

For centrifuge Z 167 M, only one rotor (220.95 V08) is acceptable. Due to this, rotor adjustments
cannot be conducted.

1.8.2 Access to the Mode: "Standard settings"

While starting this model, the following basic settings can be performed:
-
-
-
The following operating data can be retrieved in this menu:
-
-
-
-
-
-
-
-
-
-
Open the centrifuge lid and shut off the main switch. Now switch on again the main switch.For
approximately 3 seconds "Soft" will be shown in the display. Press during this time the keys "time"
(6) and "lid" (10) simultaneously. As a result, a display test is administered for approx. 3 seconds. All
possible indications will appear at the same time (see Figure 1). The LED light is flashing alternately
with green and red color.
Figure 1
ATTENTION:
-
The normal program mode can be changed back again by switching off the centrifuge,
for a short period!
All changed settings must be confirmed by the key, "start" (9). A confirmation screen
-
will appear with the word, "store", in the display "rpm/rcf" (A-1) - Only then the pre-
selections are valid!
8
Signal Turn On / Off
Keybord Sound On / Off
Sleep Mode On / Off
Number of Starts
Operating Hours of Centrifuge
Operating Hours of Motor
Software Version
Error List
Function of the Imbalance Switch
Operation of Keyboard
Hardware Version
Intermediate Circuit Voltage in Volt
Display Tests
Z167M_V1.18_eng