Attaching The Chest Electrodes; Attaching The Disposable Electrodes - EDAN SE-1200 Series User Manual

Hide thumbs Also See for SE-1200 Series:
Table of Contents

Advertisement

SE-1200 Series Electrocardiograph User Manual

3.3.2.2 Attaching the Chest Electrodes

Suction Bulb
Chest Electrode Connection:
1) Ensure that the electrodes are clean;
2) Clean the electrode area on the chest surface with 75% alcohol;
3) Daub the round area of 25mm in diameter on each electrode site with gel evenly;
4) Place a small amount of gel on the brim of the chest electrode's metal cup;
5) Place the electrode on the chest electrode site and squeeze the suction bulb.
Unclench it and the electrode is adsorbed on the chest;
6) Attach all chest electrodes in the same way.
NOTE: Long-time measurement with a strong negative pressure on the suction
bulb may cause reddening of the skin. When using the electrode on kids
or patients with delicate skin, squeeze the suction bulb lightly.

3.3.3 Attaching the Disposable Electrodes

The disposable electrodes can only be used for one time.
Limb Electrode Connection:
1) Ensure that the electrodes are clean;
2) Clean the electrode area which is a short distance
above the ankle or the wrist with 75% alcohol;
3) Daub the electrode area on the limb with gel evenly;
4) Place a small amount of gel on the metal part of the
limb electrode clamp;
5) Connect the electrode to the limb, and make sure that
the metal part is placed on the electrode area above
the ankle or the wrist;
Attach all limb electrodes in the same way.
6)
Metal Cup
Chest Electrode
CAUTION
- 27 -
Operation Preparations
Connecting to a Lead Wire

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents