A1.5 Performance Specifications - EDAN SE-601 Series User Manual

Electrocardiograph
Hide thumbs Also See for SE-601 Series:
Table of Contents

Advertisement

SE-601 Series Electrocardiograph User Manual

A1.5 Performance Specifications

Recording
Recorder:
Printing Density
Recorder Paper:
Effective Width:
Paper Speed:
HR Recognition
HR Range:
Accuracy:
ECG Unit
Leads:
Acquisition Mode:
A/D:
Resolution:
Time Constant:
Frequency Response:
Gain:
Input Impedance:
Input Circuit Current:
Input Voltage Range
Calibration Voltage:
DC Offset Voltage:
Thermal dot-matrix recorder
8 dots per mm / 200 dots per inch (amplitude axes)
40 dots per mm / 1000 dots per inch (time axes, @ 25 mm/s)
Folded thermal paper: 110mm×140mm, 150 pages
104mm
5mm/s, 6.25mm/s, 10mm/s, 12.5mm/s, 25mm/s, 50mm/s (± 3%)
30 bpm ~300 bpm
± 1 bpm
12 standard leads
Simultaneously 12 leads
24 bits
2.52uV/LSB
≥3.2s
SE-601A:
0.05Hz ~ 150Hz
SE-601B/C:
0.01Hz~300Hz
2.5, 5, 10, 20, 10/5, AGC (mm/mV), ± 5%
≥50MΩ (10Hz)
SE-601A:
≥100MΩ (10Hz)
SE-601B/C:
≤0.01μA
≤±5 mVpp
1mV± 2%
± 600mV
- 69 -