Rcd Auto - Rcd Auto Test - METREL EurotestXC MI 3152 Instruction Manual

Eurotestxc eurotestxc 2,5 kv
Table of Contents

Advertisement

MI 3152(H) EurotestXC (2,5 kV)
RCD Auto – RCD Auto test
7.8
RCD Auto test function performs a complete RCD test (trip-out time at different residual
currents, trip-out current and contact voltage) in one set of automatic tests, guided by the
instrument.
RCD Auto test procedure
RCD Auto test steps
Enter the RCD Auto function.
Set test parameters / limits.
Connect test cable to the instrument.
Connect test leads or Plug commander to the object
under test, see Figure 7.27
Start the measurement.
Test with I
, (+) positive polarity (step 1).
N
Re-activate RCD.
Test with I
, (-) negative polarity (step 2).
N
Re-activate RCD.
Test with 5I
, (+) positive polarity (step 3).
N
Re-activate RCD.
Test with 5I
, (-) negative polarity (step 4).
N
Re-activate RCD.
Test with ½I
, (+) positive polarity (step 5).
N
Test with ½I
, (-) negative polarity (step 6).
N
Trip-out current test, (+) positive polarity (step 7).
Re-activate RCD.
Trip-out current test, (-) negative polarity (step 8).
Re-activate RCD
Trip-out current test for d.c. part, (+) polarity (step 9).
Re-activate RCD
Trip-out current test for d.c. part, (-) polarity (step 10).
Re-activate RCD.
Save results (optional).
1)
Steps 9 and 10 are performed if parameter Use is set to 'other' and Type to EV RCD or
MI RCD.
Step 1
1)
.
1)
.
Notes
Start of test
RCD should trip-out
RCD should trip-out
RCD should trip-out
RCD should trip-out
RCD should not trip-out
RCD should not trip-out
RCD should trip-out
RCD should trip-out
RCD should trip-out
RCD should trip-out
End of test
108
Tests and measurements
Step 2

Advertisement

Table of Contents
loading

This manual is also suitable for:

Eurotestxc mi 3152h

Table of Contents