Rc and Rp values are indicative.
Additional error:
See Earth resistance three-wire method.

12.18 Voltage, frequency, and phase rotation

12.18.1 Phase rotation

Nominal system voltage range ............... 100 V
Nominal frequency range ....................... 14 Hz ... 500 Hz
Result displayed .................................... 1.2.3 or 3.2.1
12.18.2 Voltage
Measuring range (V)
0 ... 550
Result type ............................................ True r.m.s. (TRMS)
Nominal frequency range ....................... 0 Hz, 14 Hz ... 500 Hz
12.18.3 Frequency
Measuring range (Hz)
0.00 ... 9.99
10.0 ... 499.9
Nominal voltage range ........................... 20 V ... 550 V
12.18.4 Online terminal voltage monitor
Measuring range (V)
10 ... 550
... 550 V
AC
AC
Resolution (V)
1
Resolution (Hz)
0.01
0.1
Resolution (V)
1
186
Technical specifications
Accuracy
(2 % of reading + 2 digits)
Accuracy
(0.2 % of reading + 1 digit)
Accuracy
(2 % of reading + 2 digits)