Continuity - Continuous Resistance Measurement With Low Current - METREL EurotestXC MI 3152 Instruction Manual

Eurotestxc eurotestxc 2,5 kv
Table of Contents

Advertisement

MI 3152(H) EurotestXC (2,5 kV)
Continuity – Continuous resistance measurement with
7.6
low current
Figure 7.21: Continuity resistance measurement menu
Measurement parameters / limits
Sound
[On*, Off]
Max. resistance [Off, 0.1 Ω ... 20.0 Ω]
Limit(R)
*Instrument sounds if resistance is lower than the set limit value.
Connection diagrams
Figure 7.22: Tip commander and 3-wire test lead applications
Measurement procedure
Enter the Continuity function.
Set test parameters / limits.
Connect test cable to the instrument.
101
Tests and measurements

Advertisement

Table of Contents
loading

This manual is also suitable for:

Eurotestxc mi 3152h

Table of Contents