MI 3152(H) EurotestXC (2,5 kV)
finished single test with test results and no status
empty single test without test results
Overall statuses of Auto Sequences®
or
or
or
or
5.1.2 Structure Objects
Each Structure object has:
an icon
a name and
parameters.
Optionally they can have:
an indication of the status of the measurements under the Structure object and
a comment or a file attached.
Structure objects supported are described in Appendix D – Structure objects.
5.1.2.1 Measurement status indication under the Structure object
Overall status of measurements under each structure element /sub-element can be seen
without spreading tree menu. This feature is useful for quick evaluation of test status and as
guidance for measurements.
Figure 5.3: Structure object in tree menu
at least one single test in the Auto Sequence®
passed and no single test failed
at least one single test in the Auto Sequence®
failed
at least one single test in the Auto Sequence®
was carried out and there were no other
passed or failed single tests.
empty Auto Sequence® with empty single
tests
51
Memory Organizer