OmegaEE XD MI 3365 (25A, M, F) Instruction manual Ver. 1.2.2, Code no. 20 753 338...
Page 2
Mark on your equipment certifies that it meets requirements of all subjected EU regulations. Hereby, Metrel d.o.o. declares that the MI 3365, MI 3365 25A, MI 3365 M, and MI 3365 F are in compliance with Directive 2014/53/EU (RED) and all other subjected EU directive. The full text of the EU declaration of conformity is available at the following internet address https://www.metrel.si/DoC.
Page 3
Please note that LCD screenshots in this document may differ from the actual instrument screens in details due to firmware variations and modifications. Metrel reserve the right to make technical modifications without notice as part of the further development of the product.
MI 3365 /25A /M /F OmegaEE XD Table of Contents ABLE OF ONTENTS General description ....................10 Warnings and notes ...................... 10 1.1.1 Safety warnings ....................10 1.1.2 Warnings related to safety of measurement functions ........11 1.1.2.1 Flash HV ......................11 1.1.2.2 Differential leak., Ipe leak., Touch leak., Ileak (W-PE), Protective conductor current, Power, Leak’s &...
Page 5
MI 3365 /25A /M /F OmegaEE XD Table of Contents 4.8.10.2 Changing user password, signing out ............... 32 4.8.10.3 Managing accounts ..................33 4.8.11 Setting user permissions ..................35 4.8.12 Setting Black-box password .................. 35 Devices .......................... 36 4.10 Instrument profiles ....................... 37 4.11 Workspace Manager .....................
Page 6
MI 3365 /25A /M /F OmegaEE XD Table of Contents Selection modes ......................71 6.1.1 Single test screens ....................72 6.1.1.1 Single test start screen ..................73 6.1.1.2 Setting parameters and limits of single tests ........... 73 6.1.1.3 Single test screen during test ................74 6.1.1.4 Single test result screen ...................
Page 7
MI 3365 /25A /M /F OmegaEE XD Table of Contents 6.2.32 PE conductor (EV RCD) ..................122 6.2.32.1 Compensation of IEC plug adapter resistance ..........123 6.2.33 Functional test ....................124 ® Auto Sequences ..................... 125 ® Selection of Auto Sequences ..................
Page 8
Generic tag format...................... 166 ® Appendix D Default list of Auto Sequences ..............168 ® Appendix E Programming of Auto Sequences on Metrel ES Manager ......169 ® Auto Sequence Editor workspace ................169 ® Managing groups of Auto Sequences ............... 170 ®...
Page 9
MI 3365 /25A /M /F OmegaEE XD Table of Contents Appendix G User permissions ..................183 Standard........................183 Instructed ........................183 Appendix H OmegaEE XD models and available measuring functions ......185...
Use only standard or optional test accessories supplied by your distributor! Only test equipment provided or approved by Metrel should be connected to 3-PHASE ADAPTER connector. Do not connect external voltage to CLAMP inputs. It is intended only for connection of Clamps approved by Metrel.
Metrel Auto Sequences® are designed as guidance to tests in order to significantly reduce testing time, improve work scope and increase traceability of the tests performed. METREL assumes no responsibility for any Auto Sequence® by any means.
MI 3365 /25A /M /F OmegaEE XD General Description Dangerous high voltage is present on terminals during the test. Consider all precautions in order to avoid risk of electric shock. Mark on your equipment certifies that this equipment meets requirements of all subjected EU regulations.
MI 3365 /25A /M /F OmegaEE XD General Description Notes: A flat or faulty battery will result in an immediate power off after the instrument is disconnected from the mains. Charging time of the battery is about 14 h. 1.2.3 Power Off, Restart Instrument is powered by battery: ...
Page 14
MI 3365 /25A /M /F OmegaEE XD General Description Functionality Code of Practice Household and similar electrical appliances EN 50699 Recurrent Test of Electrical Equipment (VDE 0702) EN 50678 General procedure for verifying the effectiveness of the protective (VDE 0701) measures of electrical equipment after repair IEC/EN 62368-1 Audio/video, information and communication technology equipment - Part...
Calibration Certificate Short form instruction manual CD with instruction manual (full version) and PC SW Metrel ES Manager Standard set of the instrument MI 3365 F Instrument MI 3365 F - OmegaEE XD Bag for accessories ...
MI 3365 /25A /M /F OmegaEE XD Instrument description 3 Instrument description Front panel Figure 3.1: Left side panel Figure 3.2: Front panel Mains supply connector...
Page 17
MI 3365 /25A /M /F OmegaEE XD Instrument description F1, F2 fuses (T 16 A / 250 V) Clamp inputs Barcode scanner and RFID / NFC reader / writer device serial port PC / Printer serial port Fuse checker MicroSD card slot USB communication port Escape key / reset key Keypad...
MI 3365 /25A /M /F OmegaEE XD Instrument operation 4 Instrument operation The instrument can be manipulated via a keypad or touch screen. General meaning of keys Cursor keys are used to: select appropriate option RUN key is used to: confirm selected option start and stop measurements Escape key is used to:...
MI 3365 /25A /M /F OmegaEE XD Instrument operation Virtual keyboard Figure 4.1: Virtual keyboard Options: Toggle case between lowercase and uppercase. Active only when alphabetic characters’ keyboard layout selected. Backspace Clears last character or all characters if selected. (If held for 2 s, all characters are selected). Enter confirms new text.
MI 3365 /25A /M /F OmegaEE XD Instrument operation Presence of input PE connection, Any external voltage against earth on mains test socket Excessive leakage currents through measuring I/O’s, Too low resistance between L and N of tested device, ...
Page 21
MI 3365 /25A /M /F OmegaEE XD Instrument operation Resistance L-N < 30 Ω In pre-test a low input resistance of the device under test was measured. This can result in a high current after applying power to the device. If the high current is only of short duration (caused by a short inrush current) the test can be performed, otherwise not.
Page 22
MI 3365 /25A /M /F OmegaEE XD Instrument operation In pre-test a possible high leakage current was detected. It is likely that a dangerous leakage current (higher than 3.5 mA) will flow after applying power to the device under test. Select YES to proceed with or NO to cancel measurement.
Page 23
MI 3365 /25A /M /F OmegaEE XD Instrument operation In case of simultaneously measuring of Riso, Riso-S or Isub, Isub-S. If the voltage has dropped because of one measurement the other measurement is also compromised. Warning! A high voltage is / will be present on the instrument output! (Withstanding test voltage, Insulation test voltage, or mains voltage).
MI 3365 /25A /M /F OmegaEE XD Instrument operation Bluetooth communication active. Fuse checker Fuse checker, integrated in the OmegaEE XD instrument, is intended for verification of fuses. This functionality is active from any menu, regardless of the instrument settings, as long as the instrument is powered on.
MI 3365 /25A /M /F OmegaEE XD Instrument operation Instrument main menu From the instrument Main Menu different main operation menus can be selected. Figure 4.3: Main menu Options Single Tests Menu with single tests, see chapter 6 Single tests. Auto Sequences®...
MI 3365 /25A /M /F OmegaEE XD Instrument operation General settings In the General Settings menu general parameters and settings of the instrument can be viewed or set. Figure 4.4: Setup menu Options in General Settings menu Language Instrument language selection. Date / Time Instruments Date and time.
MI 3365 /25A /M /F OmegaEE XD Instrument operation About Basic instrument data. 4.8.1 Language In this menu the language of the instrument can be set. The selection of available languages depends on the selected instrument profile. Figure 4.5: Select language menu 4.8.2 Date and time In this menu date and time of the instrument can be set.
MI 3365 /25A /M /F OmegaEE XD Instrument operation 4.8.6 Settings Figure 4.7: Settings menu Setting options: Description Touch screen ON – touch screen is active. OFF – touch screen is deactivated. Keys & touch sound ON – sound is active. OFF –...
Page 29
MI 3365 /25A /M /F OmegaEE XD Instrument operation Result In case the measurement consists of multiple measurements taken successively, the worst or the last result can be displayed at the end. Worst – the worst result will be displayed at the end of test. Last –...
MI 3365 /25A /M /F OmegaEE XD Instrument operation 4.8.7 Bluetooth initialization In this menu the Bluetooth module is reset. Figure 4.8: Bluetooth initialization menu 4.8.8 Initial Settings In this menu internal Bluetooth module can be initialized and the instrument settings, measurement parameters and limits can be set to initial (factory) values.
MI 3365 /25A /M /F OmegaEE XD Instrument operation 4.8.9 About In this menu instrument data (name, serial number, FW (firmware) and HW (hardware) version, profile code, HD (hardware documentation) version, and date of calibration) can be viewed. Figure 4.10: Instrument info screen Note: ...
MI 3365 /25A /M /F OmegaEE XD Instrument operation Options User signing in The user should be selected first. The last used user is displayed in the first row. Goes to Password entry menu. To sign in, the selected user password must be entered and confirmed.
MI 3365 /25A /M /F OmegaEE XD Instrument operation Options Signs out the set user. Enters procedure for changing the user’s password. The actual password must be entered first followed by the new password. Enters the Account manager menu. The Account manager menu is accessed by selecting Account Manager in Sign in menu or User profile menu.
Page 34
MI 3365 /25A /M /F OmegaEE XD Instrument operation The actual password must be entered first followed by the new password. Enters menu for editing user accounts. Figure 4.14: Edit accounts menu Options Opens the window for adding a new user account. In the Add New window the name and initial password of the new user account are to be set.
MI 3365 /25A /M /F OmegaEE XD Instrument operation 4.8.11 Setting user permissions Figure 4.15: User permissions menu Options Options with different user permissions. For more information see Appendix G - User permissions 4.8.12 Setting Black-box password Black-box password can be set by administrator from the Account manager menu. Set Black-box password is valid for all users.
MI 3365 /25A /M /F OmegaEE XD Instrument operation Devices In this menu operation with external devices is configured. Figure 4.16: Devices settings menu Writing devices Type Sets appropriate writing device (Serial printer, Bluetooth printer), RFID writer). Port Sets/views communication port of selected writing device. Bluetooth device Goes to menu for pairing with selected Bluetooth device.
MI 3365 /25A /M /F OmegaEE XD Instrument operation 4.10 Instrument profiles In this menu the instrument profile can be selected from the available ones. Figure 4.17: Instrument profile menu The instrument uses different specific system and measuring settings in regard to the scope of work or country it is used.
Workspaces are stored on microSD card on directory WORKSPACES, while Exports are stored on directory EXPORTS. Export files can be read by Metrel applications that run on other devices. Exports are suitable for making backups of important works or can be used for storage of works if the removable microSD card is used as a mass storage device.
MI 3365 /25A /M /F OmegaEE XD Instrument operation List of Exports. Displays a list of Workspaces. 4.11.2.1 Operations with Workspaces Only one Workspace can be opened in the instrument at the same time. The Workspace selected in the Workspace Manager will be opened in the Memory Organizer. Figure 4.20: Workspaces menu Options Marks the opened Workspace in Memory Organizer.
MI 3365 /25A /M /F OmegaEE XD Instrument operation 4.11.2.2 Operations with Exports Figure 4.21: Workspace manager File menu Options Deletes the selected Export. Refer to chapter Deleting a Workspace / Export for more information. Imports a new Workspace from Export. Refer to chapter Importing a Workspace for more information.
MI 3365 /25A /M /F OmegaEE XD Instrument operation 4.11.2.3 Adding a new Workspace New workspaces can be added from the Workspace manager screen. Enters option for adding new Workspace. Keypad for entering name of a new Workspace is displayed after selecting New.
MI 3365 /25A /M /F OmegaEE XD Instrument operation 4.11.2.5 Deleting a Workspace / Export Workspace / Export to be deleted should be selected from the list of Workspaces / Exports. Opened workspace can’t be deleted. Enters option for deleting a Workspace / Export. Before deleting the selected Workspace / Export the user is asked for confirmation.
MI 3365 /25A /M /F OmegaEE XD Instrument operation Before the import of the selected file the user is asked for confirmation. The Imported Export file is added to the list of Workspaces. Note: If a Workspace with the same name already exists the name of the imported Workspace will be changed (name_001, name_002, name_003…).
MI 3365 /25A /M /F OmegaEE XD Instrument operation 4.12 Auto Sequence® groups The Auto Sequences® in OmegaEE XD can be organized by using lists. In a list a group of similar Auto Sequences® is stored. The Auto Sequence® groups menu is intended to manage with different lists of Auto Sequences®...
MI 3365 /25A /M /F OmegaEE XD Instrument operation 4.12.1.1 Operations in Auto Sequence® groups menu Options Opens the selected list of Auto Sequences®. Previously selected list of Auto Sequences® will be closed automatically. Refer to chapter Selecting a list of Auto Sequences® for more information. Deletes the selected list of Auto Sequences.
MI 3365 /25A /M /F OmegaEE XD Instrument operation 4.12.1.3 Deleting a list of Auto Sequences® A list of Auto Sequences® to be deleted can be selected from the Auto Sequence® groups menu. Enters option for deleting a list. Before deleting the selected list of Auto Sequences®...
MI 3365 /25A /M /F OmegaEE XD Memory Organizer 5 Memory Organizer Memory Organizer is a tool for storing and working with test data. Memory Organizer menu The data is organized in a tree structure with Structure objects and Measurements. OmegaEE XD has a multi-level structure.
MI 3365 /25A /M /F OmegaEE XD Memory Organizer Overall statuses of Auto Sequence at least one single test in the Auto sequence passed and no single test failed at least one single test in the Auto sequence failed at least one single test in the Auto sequence was carried out and there were no other passed or failed single tests empty Auto sequence with empty single...
MI 3365 /25A /M /F OmegaEE XD Memory Organizer 5.1.2.1 Measurement status indication under the Structure object Overall status of measurements under each structure element / sub-element can be seen without spreading tree menu. This feature is useful for quick evaluation of test status and as guidance for measurements.
MI 3365 /25A /M /F OmegaEE XD Memory Organizer 5.1.3 Selecting an active Workspace in Memory Organizer Memory Organizer and Workspace Manager are interconnected so an active Workspace can be selected also in the Memory Organizer menu. Procedure Press the active Workspace in Memory Organizer ...
MI 3365 /25A /M /F OmegaEE XD Memory Organizer Procedure Press the active Workspace in Memory Organizer Menu. Select Add New Structure Element in Control panel. Press “Create” to confirm. New Structure Element (Node) is added. Note: To change the name of a Node, see chapter Rename a Structure object.
Page 52
MI 3365 /25A /M /F OmegaEE XD Memory Organizer Figure 5.6: A measurement is selected in the tree menu Options Views results of measurement. The instrument goes to the measurement memory screen. Refer to chapters Single test memory screen and 7.2.4 Auto Sequence® memory screen for more information.
MI 3365 /25A /M /F OmegaEE XD Memory Organizer 5.1.5.2 Operations on Structure objects The structure object must be selected first. Figure 5.7: A structure object is selected in the tree menu Options Starts a new measurement. First type of measurement (single test or Auto sequence®) should be selected. After proper type is selected the instrument goes to single test or Auto Sequence®...
MI 3365 /25A /M /F OmegaEE XD Memory Organizer Copies & Paste a Structure object. Selected Structure object can be copied and pasted to any allowed location in structure tree. Multiple “Paste” is allowed. Refer to Chapter Copy & Paste a Structure object for more information.
Page 55
MI 3365 /25A /M /F OmegaEE XD Memory Organizer Procedure and options Select structure object to be edited. Select Parameters in Control panel. Example of Parameters menu. In menu for editing parameters the parameter’s value can be selected from a dropdown list or entered via keypad.
MI 3365 /25A /M /F OmegaEE XD Memory Organizer View or edit comments Complete comment (if exists) attached to the structure object can be seen on this screen. Press RUN key or tap on screen to open keypad for entering a new comment. 5.1.5.4 Add a new Structure Object This menu is intended to add new structure objects in the tree menu.
Page 57
MI 3365 /25A /M /F OmegaEE XD Memory Organizer Add a new structure object menu. Tap on a structure type selection window. A list of available structure items is displayed. Select one from a list. Arrow indicates the position where structure item will be inserted.
MI 3365 /25A /M /F OmegaEE XD Memory Organizer 5.1.5.5 Add a new measurement In this menu new empty measurements can be set and then added in the structure tree. The type of measurement, measurement function and its parameters are first selected and then added under the selected Structure object.
Page 59
MI 3365 /25A /M /F OmegaEE XD Memory Organizer To select another measurement, tap on filed or press the RUN to open menu for selecting measurements. Refer to Chapters 6.1 Selection modes and 7.1 Selection of Auto Sequences for more information. ...
MI 3365 /25A /M /F OmegaEE XD Memory Organizer 5.1.5.6 Clone a Structure object In this menu selected structure object can be copied (cloned) to same level in the structure tree. Cloned structure object has same name as original. Clone Figure 5.11: Clone Structure Object menu Procedure and options ...
MI 3365 /25A /M /F OmegaEE XD Memory Organizer The new structure object is displayed. 5.1.5.7 Clone a measurement By using this function, a selected empty or finished measurement can be copied (cloned) as an empty measurement to the same level in the structure tree. Procedure and options ...
MI 3365 /25A /M /F OmegaEE XD Memory Organizer 5.1.5.8 Copy & Paste a Structure object In this menu selected Structure object can be copied and pasted to any allowed location in the structure tree. Procedure and options Select the structure object to be copied. ...
MI 3365 /25A /M /F OmegaEE XD Memory Organizer The new structure object is displayed. Note: The Paste command can be executed one or more times. Options Parameters of selected structure object will be copied too. Attachments of selected structure object will be copied too.
MI 3365 /25A /M /F OmegaEE XD Memory Organizer Select Paste option from control panel. Paste The new (empty) measurement is displayed in selected Structure object. Note: The Paste command can be executed one or more times. 5.1.5.10 Cut & Paste a Structure object with sub-items In this menu selected Structure object with sub-items (sub-structures and measurements) can be cut and pasted (moved) to any allowed location in the structure tree.
MI 3365 /25A /M /F OmegaEE XD Memory Organizer 5.1.5.11 Delete a Structure object In this menu selected Structure object can be deleted. Procedure Select the structure object to be deleted. Select Delete option from control panel. Delete ...
MI 3365 /25A /M /F OmegaEE XD Memory Organizer 5.1.5.12 Delete a measurement In this menu selected measurement can be deleted. Procedure Select a measurement to be deleted. Select Delete option from control panel. Delete A confirmation window will appear. ...
MI 3365 /25A /M /F OmegaEE XD Memory Organizer Select Rename option from control panel. Rename Virtual keypad will appear on screen. Enter new text and confirm. 5.1.5.14 Recall and Retest selected measurement Procedure Select the measurement to be recalled. ...
MI 3365 /25A /M /F OmegaEE XD Memory Organizer Parameters and limits can be viewed and edited. Select Run in control panel to retest the measurement. Start Test Results / sub-results after re-run of recalled measurement. Select Save results in control panel.
Page 69
MI 3365 /25A /M /F OmegaEE XD Memory Organizer The parameters that can be searched for are displayed in the Search setup menu. Note: Equipment ID, Test date and Retest date (if applicable) refer only to the following structure objects: Appliance, Appliance FD, Medical equipment, Medical equipment FD, Welding equipment and Welding...
Page 70
MI 3365 /25A /M /F OmegaEE XD Memory Organizer Figure 5.12: Search results screen – Page view Options Next page. Previous page. Note: Search result page consist of up to 50 results. Figure 5.13: Search results screen with structure object selected Options Goes to location in Memory Organizer.
MI 3365 /25A /M /F OmegaEE XD Single tests 6 Single tests Single tests can be selected in the Main single test menu or in Memory Organizer’s main and submenus. Selection modes In Single tests main menu Area group and three different modes for selecting single tests are available.
MI 3365 /25A /M /F OmegaEE XD Single tests Cross selector This selection mode is the fastest for working with the keypad. Groups of single tests are organized in a row. For the selected group all single tests are displayed and easy accessible with up /down keys.
MI 3365 /25A /M /F OmegaEE XD Single tests 6.1.1.1 Single test start screen Single test start screen can be opened from Memory organizer or from Single test main menu. Figure 6.2: Single test start screen Options Starts the measurement. Opens help screens.
MI 3365 /25A /M /F OmegaEE XD Single tests Options Selects parameter (white) or limit (red). Selects value of parameter or limit. In case of many (multiple pages of) parameters or limits: The scroll bar on the right side of screen can be used With right / left keys it can be jumped page up / page down Enters custom value of parameter or limit.
MI 3365 /25A /M /F OmegaEE XD Single tests Aborts measurement. 6.1.1.4 Single test result screen Figure 6.5: Single test result screen Options (after measurement is finished) Starts a new measurement. Saves the result. A new measurement was selected and started from a Structure object in the structure tree: The measurement will be saved under the selected Structure object.
MI 3365 /25A /M /F OmegaEE XD Single tests Opens help screens. Refer to chapter 6.1.3 Help screens for more information. Opens screen for changing parameters and limits. Refer to chapter Setting parameters and limits of single tests for more information. Enters cross selector.
MI 3365 /25A /M /F OmegaEE XD Single tests 6.1.2 Single test (inspection) screens Visual and Functional inspections can be treated as a special class of tests. Items to be visually or functionally checked are displayed. In addition on-line statuses and other information are displayed.
MI 3365 /25A /M /F OmegaEE XD Single tests 6.1.2.2 Single test (Inspection) screen during test Figure 6.9: Inspection screen (during inspection) Options (during test) Selects item. Applies a pass to the selected item or group of items. Applies a fail to the selected item or group of items. Clears status in selected item or group of items.
MI 3365 /25A /M /F OmegaEE XD Single tests Rules for automatic applying of statuses: The parent item(s) can automatically get a status on base of statuses in child items. The fail status has highest priority. A fail status for any item will result in a fail status in all parent items and an overall fail result.
MI 3365 /25A /M /F OmegaEE XD Single tests Structure object. By pressing the key in Memory organizer menu the inspection is saved under selected location. An empty inspection was selected in structure tree and started: The result(s) will be added to the inspection. The inspection will change its status from ‘empty’...
MI 3365 /25A /M /F OmegaEE XD Single tests 6.1.3 Help screens Help screens contain diagrams for proper connection of the instrument. Figure 6.12: Examples of help screens Options: Opens help screen. Goes to previous / next help screen.
MI 3365 /25A /M /F OmegaEE XD Single tests Single test measurements 6.2.1 Visual inspection Test results / sub-results Pass, Fail, Checked Test circuit 6.2.2 Continuity // Protective earth resistance Test results / sub-results R ....Resistance Test parameters Output (Continuity) Output: [P/S –...
MI 3365 /25A /M /F OmegaEE XD Single tests Note • For the test between P/S and PE banana sockets only test current 200 mA is available. Test circuits 6.2.2.1 Compensation of test lead(s) / IEC test cable resistance Resistance of test lead(s) and cables can be compensated. Compensation is possible in following functions: •...
Page 84
MI 3365 /25A /M /F OmegaEE XD Single tests Compensation of test lead(s) / IEC test cable resistance procedure Select single test and its parameters. Connect test lead to the instrument between P/S terminal and PE terminal on test socket or short-circuit test leads connected to P/S and PE banana sockets, or connect IEC test cable between IEC connector and test socket.
MI 3365 /25A /M /F OmegaEE XD Single tests 6.2.2.2 Limit calculator Limit calculator is a tool for determining the resistance high limit. Open Limit calculator Lim. Calculator Define the limit value Set Limit rule: Length, Cross-section, Custom Limit rules: A: EN / CSA ≤...
Page 86
MI 3365 /25A /M /F OmegaEE XD Single tests ρ Specific resistance of copper 1.68×10 −8 Ωm Wire length selected from a list (1 m, 2 m, 3 m, … ,100 m) or custom numeric entry Wire cross section selected from a list (0.50 mm 2 , 0.75 mm 2 , 1.00 mm 2 , 1.50 mm 2 ,2.5 mm 2 , 4.0 mm 2 , 10.0 mm 2 ) or custom numeric entry C: NEN 3140*...
MI 3365 /25A /M /F OmegaEE XD Single tests Wire cross-section [mm L wire length R Limit [Ω] L <= 2 m 0.20 0.20 0.20 2 m < L <= 5 m 0.21 0.21 0.20 5 m < L <= 10 m 0.22 0.21 0.21...
MI 3365 /25A /M /F OmegaEE XD Single tests Test circuits Flash test 1500 V Flash test 3000 V Note • Flash test 1500 V is intended for CLASS I appliances. • Flash test 3000 V is intended for CLASS II appliances. •...
MI 3365 /25A /M /F OmegaEE XD Single tests Test circuits Riso Riso-S Riso (fixed installed DUT) Note • The current through the P/S probe is also considered in the Riso result. 6.2.5 Sub-leakage (Isub, Isub-S) Test results / sub-results Isub (+result) Sub-leakage current Result [TRMS]...
MI 3365 /25A /M /F OmegaEE XD Single tests Test circuits Isub Isub-S Isub (fixed installed DUT) Note • When P/S probe is connected during the Sub-leakage measurement, then the current through it is also considered. 6.2.6 Differential Leakage Test results / sub-results Idiff (+ mains, result) Diff: Differential Leakage current Mains [nor, rev]*...
MI 3365 /25A /M /F OmegaEE XD Single tests Delay Delay between the two steps, if Mains polarity = All, [0.1 s … 5 s] Test limits H Limit (Idiff TRMS)* H limit (Idiff, TRMS) [Off, Custom, 0.25 mA … 15.0 mA] * Result view = All - [nor, rev] are added to Test Limit names.
MI 3365 /25A /M /F OmegaEE XD Single tests Delay Delay between the two steps, if Mains polarity = All, [0.1 s … 5 s] Condition [All, NC, SFC-N, SFC-PE] NC: normal condition SFC-N: single fault, N open SFC-PE: single fault, PE open All: all tests will be carried out Test limits H Limit (Itou TRMS)*...
Page 94
MI 3365 /25A /M /F OmegaEE XD Single tests Ifi (+ result) Floating Input Leakage current Result [TRMS] no caption Actual current measurement (Itou or Ifi) [TRMS] * not shown separately in Result view = Standard Test parameters Duration Duration [Off, 2 s … 180 s] Mains polarity [All, Normal, Reversed] Normal: phase voltage is applied to the right output of...
MI 3365 /25A /M /F OmegaEE XD Single tests Test circuit Step 1 Step 2 Note • The measurement consists of three steps. Step 1 - Itou is measured. Step 2 - Ifi is measured. If there are more floating inputs, individual inputs can be measured successively and the results are added together.
Page 96
MI 3365 /25A /M /F OmegaEE XD Single tests Ifi (+ result) Floating Input leakage current Result [TRMS] no caption Actual current measurement (Ipe, Idiff or Ifi) [TRMS] * not shown separately in Result view = Standard Test parameters Duration Duration [Off, 2 s …...
MI 3365 /25A /M /F OmegaEE XD Single tests Note • The measurement consists of three steps. Step 1 - Ipe or Idiff is measured. Step 2 - Ifi is measured. If there are more floating inputs, individual inputs can be measured successively and the results are added together.
MI 3365 /25A /M /F OmegaEE XD Single tests 6.2.13 PRCD test Test results / sub-results t ∆N trip-out time t I∆N x1, (+) trip-out time (I , (+) positive polarity) N t I∆N x1, (-) trip-out time (I , (-) negative polarity) ...
MI 3365 /25A /M /F OmegaEE XD Single tests 6.2.14 RCD test Test results / sub-results t ∆N trip out time t I∆N x1, (+) trip-out time (I , (+) positive polarity) N t I∆N x1, (-) trip-out time (I , (-) negative polarity) ...
MI 3365 /25A /M /F OmegaEE XD Single tests 6.2.15 PE conductor (PRCD) Test results / sub-results R ....Resistance Result ..Indication that the protection of the varistor in PE connection works properly Test parameters Type of PRCD Design [2 pole, 3 pole, K/Di (varistor), S (3 pole), S+] Duration Duration [Off, 2 s …...
MI 3365 /25A /M /F OmegaEE XD Single tests 6.2.16 Open conductor (PRCD) In this test the instrument disconnects individual conductors on the supply side and the response of the PRCD is checked. Test results / sub-results L open Result for open L conductor [Pass, Fail] N open Result for open N conductor [Pass, Fail] PE open...
MI 3365 /25A /M /F OmegaEE XD Single tests 6.2.17 PRCD PE probe test Test results / sub-results Result Indication of the test [Pass, Fail] Test parameters / limits Test mode Test [manual, auto] Type of PRCD Design [2 pole, 3 pole, K/Di (varistor), S (3 pole), S+] Test circuits Test = manual Test = auto...
Rotation 321] Not applicable in AUS and UK versions. Wiring map parameter replaces LN cross parameter if active polarity mode is selected and Metrel 3-phase adapter (A 1322 or A 1422) is connected. Test circuits Mode = normal Mode = active...
MI 3365 /25A /M /F OmegaEE XD Single tests Note • Active polarity test is intended for testing cords equipped with (P)RCD or mains operated switches. 6.2.19 Clamp current Test results / sub-results Current Test parameters Indication of the type of Test [Differential leakage, PE leakage, Current] current measured Duration...
For more information refer to chapter Measurements according to IEC/ EN 60974-4 – Insulation resistance in 3-phase adapter user manual. Note • This test is carried out with METREL 3-phase adapter A 1422. 6.2.21 Welding Circuit Leakage – I leak (W-PE) Test results / sub-results...
For more information refer to chapter Measurements according to IEC/ EN 60974-4 – Insulation resistance in 3-phase adapter user manual. Note • This test is carried out with METREL 3-phase adapter A 1422. 6.2.22 Protective conductor current Test results / sub-results...
MI 3365 /25A /M /F OmegaEE XD Single tests • This test is carried out with METREL 3-phase adapter A 1422. • The differential current measurement principle is used for this test. 6.2.23 No-load voltage Test results / sub-results Urms...
Page 110
MI 3365 /25A /M /F OmegaEE XD Single tests Test parameters Test Test_Riso_ME [LN-PE, LN-P/S, LN-AP, PE-AP] Duration Duration [Off, 2 s … 180 s] Nominal test voltage Uiso [250 V, 500 V] Test limit L Limit (Riso) Limit(Riso) [Off, Custom, 2 MΩ … 70 MΩ] Test circuits Insulation resistance –...
MI 3365 /25A /M /F OmegaEE XD Single tests Insulation resistance – PE-AP 6.2.25 Equipment Leakage Test results / sub-results Ieq (+ mains, result) Equipment Leakage current Mains [nor, rev]* Result [TRMS, AC**, DC**] Power (only in Direct and Differential test mode) Ulpe Measured voltage between phase and earth * not shown separately in Result view = Standard...
Page 112
MI 3365 /25A /M /F OmegaEE XD Single tests * Result view = All - [nor, rev] are added to Test Limit names. Test limit remains the same. Test circuits Equipment Leakage (alternative) on Class I ME Equipment Leakage (alternative) on Class II ME Equipment Leakage (direct) on Class I...
MI 3365 /25A /M /F OmegaEE XD Single tests Equipment Leakage (differential) on Class I ME Equipment Leakage (direct, differential) on Class II ME 6.2.26 Applied Part Leakage Test results / sub-results Iap (+ mains, Vext, result) Applied Part Leakage current Mains voltage [nor, rev]* Vext on AP [nor, rev]* Result [TRMS]...
Page 114
MI 3365 /25A /M /F OmegaEE XD Single tests Delay Delay between the two steps, if Mains polarity = All, [0.1 s … 5 s] Vext AP [All, Normal, Reversed] Normal: Vext on AP has same phase as mains voltage. Reversed: Vext on AP has opposite phase as mains voltage.
MI 3365 /25A /M /F OmegaEE XD Single tests Applied part leakage (direct) on Class I ME Applied part leakage (direct) on Class II ME 6.2.27 Touch Current (medical equipment) Test results / sub-results Itou (+ mains, condition, result) Touch current Mains [nor, rev]* Result [TRMS, AC, DC] Power...
MI 3365 /25A /M /F OmegaEE XD Single tests Condition [All, NC, SFC-N, SFC-PE] NC: normal condition SFC-N: single fault, N open SFC-PE: single fault, PE open All: all tests will be carried out Unom Nominal line to earth voltage whose value corresponds to the nominal mains voltage Unom [100 V …...
Page 117
MI 3365 /25A /M /F OmegaEE XD Single tests Normal: phase voltage is applied to the right output of the mains test socket. Reversed: phase voltage is applied to the left output of the mains test socket. All: tests at both mains polarities will be carried out. Delay Delay between the two steps, if Mains polarity = All, [0.1 s …...
MI 3365 /25A /M /F OmegaEE XD Single tests 6.2.29 SELV/PELV Voltage Test results / sub-results U (+ mains, condition, result) Voltage Mains [nor, rev]* Result [TRMS, AC, DC] * not shown separately in Result view = Standard Test parameters Duration Duration [Off, 2 s …...
MI 3365 /25A /M /F OmegaEE XD Single tests Test circuit SELV / PELV Note • Voltmeter is floating against PE. 6.2.30 EVSE Diagnostic Test (A 1632) Test results / sub-results Maximal value of CP (control pilot) signal Minimal value of CP (control pilot) signal Duty cycle of CP (control pilot) signal Freq Frequency of CP (control pilot) signal...
Page 120
[Remote (Bluetooth), Manual (A 1632)] Test circuits Notes • Refer to A 1632 eMobility Analyser Instruction manual for more information • Refer to application note Guide to OmegaPAT/GT XA for more information • Follow instructions in Metrel Auto Sequences® for charging cables...
MI 3365 /25A /M /F OmegaEE XD Single tests 6.2.31 EV-RCD Test results / sub-results t ∆N trip out time t I∆N (+) trip-out time at (+) positive polarity t I∆N (-) trip-out time at (-) negative polarity I ∆N trip-out current I I∆N (+) trip-out current at (+) positive polarity...
• Refer to A 1832 eMobility Adapter Instruction manual for more information. • Refer to application note Guide to OmegaPAT/GT XA for more information. • Follow instructions in Metrel Auto Sequences® for charging cables. 6.2.32 PE conductor (EV RCD) Test results / sub-results R ....
MI 3365 /25A /M /F OmegaEE XD Single tests Notes • PE resistance of IEC plug adapter can be compensated. See chapter Compensation of IEC plug adapter resistance for details. 6.2.32.1 Compensation of IEC plug adapter resistance Resistance of IEC plug adapter can be compensated in PE conductor (EV RCD) test function: Connection for compensating the resistance of IEC plug adapter Compensation of IEC plug adapter resistance procedure Select PE conductor (EV RCD) single test and its parameters.
MI 3365 /25A /M /F OmegaEE XD Single tests Notes • It is recommended to compensate leads resistance with the same test current as the measurement will be made. 6.2.33 Functional test Test results / sub-results Pass, Fail, Checked Specific option Power is applied to the mains test socket to power up the tested equipment during a functional inspection.
Auto sequence® can be stored in the memory together with all related information. Auto Sequences® can be pre-programmed on PC with the Metrel ES Manager software and uploaded to the instrument. On the instrument parameters and limits of individual single test in the Auto Sequence®...
MI 3365 /25A /M /F OmegaEE XD Auto Sequences® New Auto Sequence® group is selected and all Auto Sequences® within that group are displayed on the screen. ® 7.1.2 Searching in Auto Sequences menu In Auto Sequence® menu it is possible to search for Auto Sequences® on base of their Name or Short code.
Page 127
MI 3365 /25A /M /F OmegaEE XD Auto Sequences® Clears all filters. Sets filters to default value. Clear filters Searches through the active Auto Sequence® group according filters. Search The results are shown in the Search results screen presented on Figure 7.1 and Figure 7.2. Figure 7.1: Search results screen –...
MI 3365 /25A /M /F OmegaEE XD Auto Sequences® Goes to Auto Sequence® view menu. Starts the selected Auto Sequence®. ® ® 7.1.3 Organization of Auto Sequences in Auto Sequences menu The Auto Sequences® to be carried out can be selected from the Main Auto Sequences® menu. This menu can be organized in a structural manner with folders, sub-folders and Auto Sequences®.
MI 3365 /25A /M /F OmegaEE XD Auto Sequences® ® Organization of an Auto Sequence An Auto Sequence® is divided into three phases: Before starting the first test the Auto Sequence® view menu is shown (unless it was started directly from the Main Auto Sequences®...
MI 3365 /25A /M /F OmegaEE XD Auto Sequences® ® 7.2.1.2 Auto Sequence view menu (measurement is selected) Figure 7.5: Auto Sequence® view menu – measurement selected Options: Selects single test. Opens menu for changing parameters and limits of selected measurements.
MI 3365 /25A /M /F OmegaEE XD Auto Sequences® Selects Multiple points. Sets operation mode for multiple points. For more information see chapter Managing multiple points. 7.2.1.3 Auto Sequence® Configurator menu Auto Sequence® Configurator menu options are active only when single tests within selected Auto sequence®...
MI 3365 /25A /M /F OmegaEE XD Auto Sequences® Options: Starts the Auto Sequence® from Configurator menu. Option is available, if configurator was opened from Auto Sequence® Main menu. New configuration is automatically applied to all related single tests. Confirms limits and parameters setting and returns to view menu. Option is available, if configurator was opened from Auto Sequence®...
Page 133
MI 3365 /25A /M /F OmegaEE XD Auto Sequences® Figure 7.7: Auto Sequence® – Example of a pause with message Figure 7.8: Auto Sequence® – Example of a finished measurement with options for proceeding Options (during execution of an Auto Sequence®): Proceeds to the next step in the test sequence.
For more information about programming Auto Sequences® refer to chapter Appendix E Programming of Auto Sequences® on Metrel ES Manager. If Inspection Expert mode flow command is set, the visual inspection screen and Functional inspection screen are displayed for 1 second and an overall PASS is automatically applied at the end of test.
Page 135
MI 3365 /25A /M /F OmegaEE XD Auto Sequences® View results of individual measurements. The instrument goes to menu for viewing details of the Auto Sequence®. Saves the Auto Sequence® results. A new Auto Sequence® was selected and started from a Structure object in the structure tree: ...
MI 3365 /25A /M /F OmegaEE XD Auto Sequences® Note: Options menu content depends on Devices settings menu. If no writing device is set, then ‘Print label’ and ‘Write RFID’ icons are hidden. Only one writing device can be set at the same time.
Page 137
MI 3365 /25A /M /F OmegaEE XD Auto Sequences® Enters menu for viewing details of the Auto Sequence®. Refer to chapter 7.2.3 Auto Sequence® result screen for more information. Print label. For more information refer to chapter Appendix C - Print labels and write / read RFID / NFC tags ®...
MI 3365 /25A /M /F OmegaEE XD Maintenance 8 Maintenance Periodic calibration It is essential that all measuring instruments are regularly calibrated in order for the technical specification listed in this manual to be guaranteed. We recommend an annual calibration. Fuses There are two fuses on the left side panel: F1, F2: T 16 A / 250 V / (32 ...
MI 3365 /25A /M /F OmegaEE XD Communications 9 Communications The instrument can communicate with the Metrel ES Manager PC software. The following action is supported: Saved results and Tree structure from Memory organizer can be downloaded and stored to a PC.
It is possible to communicate with serial scanners and RFID / NFC reader / writer device via the BARCODE serial port and with serial printers via the PC / PRINTER serial port. Contact Metrel or your distributor which external devices and functionalities are supported. See Chapter 4.9...
See 3-phase adapter Instruction manual for more details. Do not connect other devices than Metrel test adapters to the 3 PHASE ADAPTER connector. 9.5.2 110 V Test Adapter (A 1474) 110 V Adapter (A 1474) is used as interface to connect 110 V appliances to the instrument.
MI 3365 /25A /M /F OmegaEE XD Communications 9.5.3 Tip Commander (A 1694) Tip Commander (A 1694) enables remote controlled execution of measurements. Figure 9.3: Connecting Tip Commander (A 1694) Note: Tip Commander (A 1694) should be enabled in Settings menu for operation.
MI 3365 /25A /M /F OmegaEE XD Technical specifications Output voltage Range Resolution Accuracy (3 % of reading + 2 D) 0 V … 600 V Nominal voltages Un ........250 V, 500 V (- 0 %, + 10 %) Short circuit current ........
MI 3365 /25A /M /F OmegaEE XD Technical specifications Influence of load current ........< 0.02 mA/A Test terminals: Idiff Socket L,N – Socket PE, P/S Socket L – Socket N 10.5 Touch Leakage Touch leakage current Range Resolution Accuracy (3 % of reading + 3 D) Itou 0.000 mA …...
MI 3365 /25A /M /F OmegaEE XD Technical specifications Test terminals: Socket L,N – Socket PE Socket L – Socket N 10.7 Ipe+Ifloating input (Ipe+Ifi) PE leakage current Range Resolution Accuracy (3 % of reading + 3 D) 0.000 mA … 1.999 mA 1 µA ...
MI 3365 /25A /M /F OmegaEE XD Technical specifications Range Resolution Accuracy (3 % of reading + 3 D) 0.000 mA … 1.999 mA 1 µA 5 % of reading 2.00 mA … 19.99 mA 0.01 mA Operating range (acc. to EN 61557-16) .... 0.020 mA … 19.99 mA Output voltage ..........
MI 3365 /25A /M /F OmegaEE XD Technical specifications Total Harmonic Distortion (voltage) Range Resolution Accuracy (5 % of reading + 5 D) THDU 0.0 % … 99.9 % 0.1 % Total Harmonic Distortion (current) Range Resolution Accuracy (5 % of reading + 5 D) 0 mA …...
Page 149
MI 3365 /25A /M /F OmegaEE XD Technical specifications Power (active) Range Resolution Accuracy (5 % of reading + 5 D) 0.00 W … 19.99 W 0.01 W 5 % of reading 20.0 W … 199.9 W 0.1 W ...
MI 3365 /25A /M /F OmegaEE XD Technical specifications Current Range Resolution Accuracy (3 % of reading + 5 D) 0 mA … 999 mA 1 mA 3 % of reading 1.00 A … 16.00 A 0.01 A Voltage Range Resolution Accuracy...
MI 3365 /25A /M /F OmegaEE XD Technical specifications 10.12 RCD test Trip-out time Range Resolution Accuracy 3 ms 1 ms 0 ms … 300 ms (999 ms*) (½I N 3 ms tΔN 1 ms 0 ms … 300 ms (I ...
MI 3365 /25A /M /F OmegaEE XD Technical specifications Test terminals: Socket PE – IEC PE PE conductor (Type = K/ Di (varistor)) Test principle: A voltage is applied between PE connections of the PRCD-K. There is a ‘PASS’ if PRCD trips. Open circuit voltage .....
MI 3365 /25A /M /F OmegaEE XD Technical specifications 10.16 EV RCD test Trip-out time Result Test current Range Resolution Accuracy ½ I 3 ms 0.0 ms … 300.0 ms 0.1 ms N 3 ms 0.0 ms … 300.0 ms 0.1 ms a.c.
MI 3365 /25A /M /F OmegaEE XD Technical specifications Current source (at nominal mains voltage, use of standard accessories) ................> 0.2 A at R < 2 Ω Open circuit voltage .......... < 9 V a.c. PE conductor (I test = Low) Range Resolution Accuracy...
MI 3365 /25A /M /F OmegaEE XD Technical specifications 10.21 Flash test Current a.c. (apparent) Range Resolution Accuracy (5 % of reading + 5 D) 0.00 mA … 2.50 mA 0.01 mA Open circuit test voltage ....1500 V, 3000 V (-0/+5%) at supply voltage 115 V, 230 V Short circuit current………………………….<...
MI 3365 /25A /M /F OmegaEE XD Technical specifications 10.23 Welding Circuit leakage (Ileak W-PE) Refer to chapter Technical specifications in 3-phase adapter instrument user manual. 10.24 Protective Conductor current (I diff) Refer to chapter Technical specifications in 3-phase adapter instrument user manual. 10.25 No-load voltage Refer to chapter Technical specifications in 3-phase adapter instrument user manual.
MI 3365 /25A /M /F OmegaEE XD Technical specifications Operating range direct and differential method (acc. to EN 61557-16)…..0.010 mA … 19.99 mA Operating range alternative method (acc. to EN 61557-16)……………….0.020 mA … 19.99 mA Ulpe (direct, differential, alternative) Range Resolution Accuracy (2 % of reading + 2 D)
MI 3365 /25A /M /F OmegaEE XD Technical specifications Test terminals: Iap (alternative) Socket LNPE, PE – P/AP Iap (direct) Socket L,N,PE, PE – P/AP Socket PE, PE – P/AP Socket L – Socket N 10.29 Touch Current (medical equipment) Touch current Range Resolution...
MI 3365 /25A /M /F OmegaEE XD Technical specifications Ulpe Range Resolution Accuracy (2 % of reading + 2 D) 0 V … 299 V Ulpe Test terminals Socket L,N - P/S Ulpe Socket L – Socket PE Socket L – Socket N 10.31 SELV/PELV Voltage Voltage (U trms, U ac) Range...
MI 3365 /25A /M /F OmegaEE XD Technical specifications 10.32 Fuse checker Test voltage..............5 Vdc Test current ..............< 1 mA Result ................Continuous sound – Fuse OK No sound – Fuse blown Test terminals Fuse Fuse checker pins 10.33 General data Mains supply Supply voltage, frequency .......
Page 161
MI 3365 /25A /M /F OmegaEE XD Technical specifications Bluetooth ..........Class 1 Dimensions (w×h×d): ....... 31 cm × 13 cm × 25 cm Weight ............. 6.1 kg Reference conditions Reference temperature range: ....15 C … 35 C Reference humidity range: ...... 35 % … 65 % RH Operation conditions Working temperature range: ....
MI 3365 /25A /M /F OmegaEE XD Appendix A Appendix A Structure objects in OmegaEE XD Structure elements used in Memory Organizer are instrument’s Profile dependent. Symbol Default name Description Node Node Project Project Location Location Client Client Appliance Appliance (basic description) Appliance FD Appliance (full description) Medical device...
– barcode or QR code – in addition. The instrument supports RFID / NFC reader / writer device; tag type supported is NTAG216. Please check with Metrel or distributor which printers and labels are supported in your OmegaEE XD instrument.
Page 165
MI 3365 /25A /M /F OmegaEE XD Appendix C Label type Form Tag content Data1 label Data 2 label size arrangement W x H (mm) Classic Barcode Test code, appliance ID Appliance ID Text Test code, appliance ID, test Appliance ID, test or or retest date, status, user retest date, status, user Test code, appliance ID,...
MI 3365 /25A /M /F OmegaEE XD Appendix C Notes: label is intended to mark supply cords. Data not available will not be printed on the label. Test or Retest date: is set in the General Settings => Devices => Writing devices menu. ...
Page 167
MI 3365 /25A /M /F OmegaEE XD Appendix C Form size Tag content Data W x H arrangement (mm) Text Parent object name, Test code, Object ID, test or retest date, status, user 50 x 25.5 Parent object name, Test code, Object ID, test date, retest period, Auto Sequence®...
MI 3365 /25A /M /F OmegaEE XD Appendix D ® Appendix D Default list of Auto Sequences Default list of Auto Sequences® for OmegaEE XD instrument is available on Metrel home page: https://www.metrel.si...
Appendix E Programming of Auto Sequences Metrel ES Manager The Auto Sequence® editor is a part of the Metrel ES Manager software. In Auto Sequence® editor Auto Sequence® can be pre-programmed and organized in groups, before uploaded to the instrument.
MI 3365 /25A /M /F OmegaEE XD Appendix E appropriate Single tests (measurements, inspections and custom inspections) and Flow commands and setting their parameters, arbitrary Auto Sequences® can be created. Figure E.2: Example of an Auto Sequence® header Figure E.3: Example of a measurement step Figure E.4: Example of an Auto Sequence®...
Page 171
MI 3365 /25A /M /F OmegaEE XD Appendix E Operation options on Group of Auto Sequences® are available from menu bar at the top of Auto Sequence® Editor workspace. File operation options: Opens a file (Group of Auto Sequences®). Creates a new file (Group of Auto Sequences®). Saves / Saves as the opened Group of Auto Sequences®...
MI 3365 /25A /M /F OmegaEE XD Appendix E Auto Sequence®: Copy to clipboard Folder: Copy to clipboard including subfolders and Auto Sequences® Auto Sequence®: Paste it to selected location Folder: Paste it to selected location Auto Sequence®: Creates shortcut to selected Auto Sequence® Double click on the object name allows it name edit: Auto Sequence®...
MI 3365 /25A /M /F OmegaEE XD Appendix E ® E.2.2 Search within selected Auto Sequence group By entering the text into search box and click on the search icon, found results are ® highlighted with yellow background and first found result (Folder or Auto Sequence ) is focused.
Result step. E.3.2 Single tests Single tests are the same as in Metrel ES Manager Measurement menu. Limits and parameters of the measurements can be set. Results and sub-results can’t be set. E.3.3 Flow commands Flow commands are used to control the flow of measurements.
MI 3365 /25A /M /F OmegaEE XD Appendix E Right click on the selected measurement step / flow command: Copy – Paste before A measurement step / flow command can be copied and pasted above selected location on the same or on another Auto Sequence®.
Page 176
MI 3365 /25A /M /F OmegaEE XD Appendix E No notifications mode Instrument skips pre-test warnings (see chapter 4.5 Symbols and messages for more information). Parameters: State On – enables No notifications mode Off – disables No notifications mode Appliance info Instrument enables to automatically select the appliance type and add the Appliance ID, Appliance name and Retest period to the Auto Sequence®.
Page 177
MI 3365 /25A /M /F OmegaEE XD Appendix E Parameters: State On – enables automatic settings of tickers in Visual and Functional tests. Off – disables automatic settings of tickers in Visual and Functional tests. Operation after end of test This flow command controls the proceeding of the Auto Sequence®...
MI 3365 /25A /M /F OmegaEE XD Appendix E E.6 Custom Inspections programming Arbitrary set of tasks dedicated to specific user defined Inspections can be programmed with application of Custom Inspection Editor Tool, accessible from Auto Sequence® Editor workspace. Custom Inspections are stored in dedicated file *.indf with user defined name. For application of Custom Inspections as a single test within Auto Sequence®...
Page 179
MI 3365 /25A /M /F OmegaEE XD Appendix E Saves / Saves as Custom Inspection Data file opened on active tab. Menu for browsing to the folder location and editing of file name is opened. Browse to the location, confirm overwriting, if file already exists or edit file name to save it as a new Custom Inspection Data file.
E.6.2 Applying Custom Inspections Custom inspections can be applied in Auto Sequences®. Direct assignment of Custom inspection to the Metrel ES manager structure objects is not possible. After custom created Inspection Data file is opened, available inspections are listed in Custom Inspections tab of Single test area of Auto Sequence®...
Page 181
After confirmation of selection, new Inspection Data file is opened and list of available Custom Inspections is changed. Note: If Metrel ES Manager Work scope is changed, opened Inspection Data file remains active and available Custom Inspections remains the same.
MI 3365 /25A /M /F OmegaEE XD Appendix F Appendix F Testing in IT or CT supply systems When testing appliances in IT or CT supply system, some test functions in the instrument are omitted due to safety reasons. Following is a list of applicable test functions. Supply system Test function ✓...
Page 183
Auto equence®can be used: • if the ‘Instructed person use’ checkbox is checked. This checkbox can be set in Metrel ES Manager, in Auto Sequence® editor. In general, changing of measuring parameters and limits on included single tests are not possible.
Page 184
MI 3365 /25A /M /F OmegaEE XD Appendix G Settings - Limitations Test mode Limited to standard Auto Sequence® flow Limited to ends if fail Result type Limited to worst Profiles The Profiles menu is not accessible. Initial settings Initial settings cannot be carried out.
Page 185
MI 3365 /25A /M /F OmegaEE XD Appendix H Appendix H OmegaEE XD models and available measuring functions Measuring function MI 3365 MI 3365 MI 3365 MI 3365 Visual inspections • • • • Fuse test • • • • Continuity // Protective earth resistance 200 mA •...
Page 186
MI 3365 /25A /M /F OmegaEE XD Appendix H Note Measurements can only be performed in combination with optional accessories: • IEC/EN 60974-4 measurements are supported with active 3-phase adapter A 1422 • Clamp current leakage measurements are supported with optional clamps A 1579...