METREL EurotestXC MI 3152 Instruction Manual page 191

Eurotestxc eurotestxc 2,5 kv
Table of Contents

Advertisement

MI 3152(H) EurotestXC (2,5 kV)
12.25 Illumination
Illumination (A 1172)
Specified accuracy is valid for complete operating range.
Measuring range (lux)
0.01 ... 19.99
20.0 ... 199.9
200 ... 1999
2.00 ... 19.99 k
Measurement principle ........................... silicon photodiode with V() filter
Spectral response error .......................... < 3.8 % according to CIE curve
Cosine error ............................................ < 2.5 % up to an incident angle of  85
Overall accuracy ..................................... matched to DIN 5032 class B standard
Illumination (A 1173)
Specified accuracy is valid for complete operating range.
Measuring range (lux)
0.01 ... 19.99
20.0 ... 199.9
200 ... 1999
2.00 ... 19.99 k
Measurement principle ........................... silicon photodiode
Cosine error ............................................ < 2.5 % up to an incident angle of  85
Overall accuracy ..................................... matched to DIN 5032 class C standard
12.26 Auto Sequences®
Refer to each individual test (measurement) for detailed technical specification.
Resolution (lux)
0.01
0.1
1
10
Resolution (lux)
0.01
0.1
1
10
190
Technical specifications
Accuracy
(5 % of reading + 2 digits)
(5 % of reading)
O
Accuracy
(10 % of reading + 3 digits)
(10 % of reading)
O